Maintenance & Troubleshooting
Figure 5-10 – Installation of
Electronics Assembly Connector Plug
5. Push the electronics assembly (4) into the electronics
housing (5) and fasten with the two 6-32 captive
screws. Torque each captive screw to 7in-lbs ±1 in-
lbs (0.8 N-m ±0.1 N-m), or hand-tight (see
Figure 5-11 – Installation of
Electronics Assembly
Terminal Block Assembly
1. Install the terminal block assembly (6) into the “FIELD
TERMINALS” side of the electronics housing (5) and
torque the two 6-32 screws to 7in-lbs ± 1 in-lbs (0.8 N-
m ±0.1 N-m), or hand-tight.