4.4.4
Setting the correct bottom reference (on the 3102/3105)
Screen display:
b.rEF
Default:
11
The transmitter leaves the factory with the bottom reference pre-programmed to the
maximum measurement range of 36 ft. (11 m).
Figure 4-3. Transmitter Bottom Reference
A. Transmitter Bottom Reference
B. Distance-to-target
C. Depth (Level)
To change the bottom reference
1.
If entering the menu system from the PV display, press the green button
repeatedly
until the “b.rEF” menu option is indicated.
2.
Press the blue button
to enter the “b.rEF” menu and display the present bottom
reference (b.rEF) value.
3.
If this value is correct, press the red button
and then the green button
to get to the
next menu option. Otherwise, continue with step (
4.
Press the blue button
to start the editing mode. The first digit flashes to indicate it
can now be edited.
5.
Press the green button
repeatedly to edit the flashing digit.
6.
Press the blue button
to select the next digit. The digit flashes to indicate it can be
edited.
7.
Repeat steps (
) until the last digit is flashing, and edited as required.
8.
Press the blue button
to confirm the new b.rEF value. None of the digits should now
be flashing.
B
C
A
Summary of Contents for Rosemount 3101
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 4840 Rev CB Title Page February 2015 ...
Page 12: ...x Reference Manual 00809 0100 4840 Rev CB Contents February 2015 Contents ...
Page 34: ...22 Reference Manual 00809 0100 4840 Rev CB Section 3 Installation February 2015 Installation ...
Page 68: ...56 Reference Manual 00809 0100 4840 Rev CB Section 4 Starting up February 2015 Starting up ...
Page 236: ...224 Reference Manual 00809 0100 4840 Rev CB Index February 2015 Index ...
Page 237: ......
Page 238: ......
Page 239: ......