Wiring the ac and dc voltage versions
Power supply (ac version):
•
24, 48, 115 , or 230 Vac (50/60 Hz), maximum of 4 VA
•
External fuse: maximum 10 A, fast or slow, HBC, 250 Vac
Note
The supply voltage is selected when ordering the level switch.
All voltages are ±10% (EN 61010).
Power supply (dc version):
•
24 Vdc ±15%, maximum of 2.5 W
•
External fuse: not required
Note
The supply voltage is selected when ordering the level switch.
The voltage variance of ±15% includes the ±10% of EN 61010.
Signal output (ac and dc versions):
•
Micro-switched, SPDT relay contacts
•
Maximum 250 Vac, 5 A, non-inductive
•
Maximum 30 Vdc, 4 A, non-inductive
Figure 3-2: Terminal Connections (Ac and Dc Voltage Versions)
5
6
7
2
1
L N
+
PE
-
B
A
A. Signal output connections
B. Power supply connections
Maximum wire size is 4 mm
2
(AWG12).
October 2019
Quick Start Guide
Quick Start Guide
23