Emerson Rosemount 226 Instruction Manual Download Page 2

CAUTION

SENSOR/PROCESS

APPLICATION COMPATIBILITY

The wetted sensor materials may not be
compatible with process com 

position

and operating conditions. Application
compat ibility is entirely the responsibili-
ty of the user.

DANGER

HAZARDOUS AREA INSTALLATION

Installations near flammable liquids or in hazardous
area locations must be carefully evaluated by qual-
ified on site safety personnel. This sensor is not
Intrinsically Safe or Explosion Proof. 

To secure and maintain an intrinsically safe instal-
lation, the certified safety barrier, transmitter, and
sensor combi nation must be used. The installa-
tion system must comply with the governing
approval  agency (FM, CSA or BASEEFA/CEN-
ELEC)  hazardous area classification require-
ments. Consult your analyzer/transmitter instruc -
tion manual for details.

Proper installation, operation and servicing of this
sensor in a Hazardous Area Instal lation is entirely
the responsibility of the user.

ESSENTIAL  INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!

Rosemount Analytical designs, manufactures, and tests its products to meet
many national and international standards. Because these instruments are
sophisticated technical products, you must properly install, use, and main-
tain them to ensure they continue to operate within their normal specifica-
tions. The following instructions must be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount
Analytical products. Failure to follow the proper instructions may cause any
one of the following situations to occur: Loss of life; personal injury; proper-
ty damage; damage to this instrument; and warranty invalidation.

• Read all instructions prior to installing, operating, and servicing the prod-

uct. If this Instruction Manual is not the correct manual, telephone 1-800-
654-7768 and the requested manual will be provided. Save this
Instruction Manual for future reference.

• If you do not understand any of the instructions, contact your Rosemount

representative for clarification.

• Follow all warnings, cautions, and instructions marked on and supplied

with the product.

• Inform and educate your personnel in the proper installation, operation,

and maintenance of the product.

• Install your equipment as specified in the Installation Instructions of the

appropriate Instruction Manual and per applicable local and national
codes. Connect all products to the proper electrical and pressure sources.

• To ensure proper performance, use qualified personnel to install, operate,

update, program, and maintain the product.

• When replacement parts are required, ensure that qualified people use

replacement parts specified by Rosemount. Unauthorized parts and pro-
cedures can affect the product’s performance and place the safe opera-
tion of your process at risk. Look alike substitutions may result in fire, elec-
trical hazards, or improper operation.

• Ensure that all equipment doors are closed and protective covers are in

place, except when maintenance is being performed by qualified persons,
to prevent electrical shock and personal injury.

Emerson Process Management

Liquid Division

2400 Barranca Parkway
Irvine, CA 92606  USA
Tel: (949) 757-8500
Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2003

About This Document

This manual contains instructions for installation and operation of the Model 226 Submersion/Insertion
Toroidal Conductivity Sensor. The following list provides notes concerning all revisions of this document.

Rev. Level

Date

Notes

0

5/01

This is the initial release of the product manual. The manual has been refor-
matted to reflect the Emerson documentation style and updated to reflect any 
changes in the product offering.

A

3/02

Updated multiple drawings throughout.

B

3/12

Updated page 18.

Summary of Contents for Rosemount 226

Page 1: ...Model 226 Submersion Insertion Sensor Instruction Manual PN 51 226 rev B March 2012 ...

Page 2: ...e correct manual telephone 1 800 654 7768 and the requested manual will be provided Save this Instruction Manual for future reference If you do not understand any of the instructions contact your Rosemount representative for clarification Follow all warnings cautions and instructions marked on and supplied with the product Inform and educate your personnel in the proper installation operation and ...

Page 3: ...pecifications 1 1 3 Ordering Information 2 2 0 INSTALLATION AND WIRING 3 2 1 Sensor Installation 3 2 2 Sensor Wiring 3 2 3 Extension Cable and Junction Box Use 3 3 0 CALIBRATION 13 4 0 MAINTENANCE AND TROUBLESHOOTING 14 4 1 Sensor Maintenance 14 4 2 Troubleshooting 14 5 0 RETURN OF MATERIAL 16 5 1 General 16 5 2 Warranty Repair 16 5 3 Non Warranty Repair 16 i ...

Page 4: ... Extension 8 2 6 Extension Cable PN 9200245 Instrument End Model 1181T 9 2 7 Extension Cable PN 9200245 Instrument End Model 1054 and 2054 series 10 2 8 Extension Cable PN 9200245 Instrument End Model 54C and 2081T 11 2 9 Wiring Model 226 56 Sensor to Model 1055 01 21 31 Analyzer 12 2 10 Wiring Model 226 54 Sensor to Model 1055 21 31 Analyzer 12 LIST OF TABLES Table No Title Page 1 1 Model 226 Acc...

Page 5: ...s the threaded mounting shaft This subassembly is then molded in chemically resistant plastic There are no seams or welds that can crack and cause leakage and subsequent failure 1 2 MODEL 226 SPECIFICATIONS Minimum Range w 1181T 0 to 80 µS cm Code 50 Maximum Range w 1181T 0 to 160 mS cm Code 50 Minimum Range w 1054 2054 Series 0 to 50 µS cm Maximum Range w 54C 1054B 2081T 81T 3081T 4081T 1054 1054...

Page 6: ...cement Parts TABLE 1 2 Prepped Cable Extension Chart for Model 226 MODEL 226 LARGE BORE TOROIDAL CONDUCTIVITY SENSOR CODE SENSOR CONSTRUCTION Required Selection 02 PEEK standard temperature to 120 C 248 F Note 1 226 02 50 80 EXAMPLE NOTES 1 Cables can be extended with the use of remote junction box PN 23550 00 see Table 1 2 for interconnect cable 2 The sensor is supplied with an EPDM gasket A Vito...

Page 7: ...o be mounted through any user supplied flange However the user is responsible for cutting a hole through the flange to fit the Model 226 The sensor has a 7 8 in 9 UNC thread so the flange may be drilled and tapped for this thread Alternatively a simple 15 16 in 2 4 cm drilled hole will accommo date the 7 8 in 9 thread of the sensor 1 A registered trademark of E I du Pont de Nemours Co 2 2 SENSOR W...

Page 8: ...4 MODEL 226 SECTION 2 0 INSTALLATION AND WIRING FIGURE 2 1 Dimensional Diagrams WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH DWG NO REV 40022604 A SUBMERSION INSERTION DWG NO REV 40022607 A ...

Page 9: ...5 FIGURE 2 2 Model 226 Sensor Wiring Directly to Instrument MODEL 226 SECTION 2 0 INSTALLATION AND WIRING WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH ...

Page 10: ... 226 SECTION 2 0 INSTALLATION AND WIRING FIGURE 2 3 Model 226 Toroidal Conductivity Sensor Wiring With Remote Junction Box PN 23550 00 DWG NO REV 40022819 G WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH ...

Page 11: ...226 SECTION 2 0 INSTALLATION AND WIRING FIGURE 2 4 Model 226 55 Toroidal Conductivity Sensor Wiring With Remote Junction Box PN 23293 00 DWG NO REV 40022806 R WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH ...

Page 12: ...repped Cable Extension MILLIMETER INCH DWG NO REV 40022511 B Unprepped Cable If unprepped extension cable PN 9200245 is selected refer to Figures 2 5 through 2 8 on the following pages for detailed instructions on cable end preparation for each sensor to transmitter analyzer configuration ...

Page 13: ...9 MODEL 226 SECTION 2 0 INSTALLATION AND WIRING FIGURE 2 6 Extension Cable PN 9200245 Instrument End Model 1181T DWG NO REV 40118142 B ...

Page 14: ...10 MODEL 226 SECTION 2 0 INSTALLATION AND WIRING FIGURE 2 7 Extension Cable PN 9200245 Instrument End Model 1054 and 2054 series DWG NO REV 40205413 B ...

Page 15: ...11 MODEL 226 SECTION 2 0 INSTALLATION AND WIRING FIGURE 2 8 Extension Cable PN 9200245 Instrument End Model 54C and 2081T DWG NO REV 40022513 B ...

Page 16: ...12 FIGURE 2 9 Wiring Model 226 56 Sensor to Model 1055 01 21 31 Analyzer FIGURE 2 10 Wiring Model 226 54 Sensor to Model 1055 21 31 Analyzer MODEL 226 SECTION 2 0 INSTALLATION AND WIRING ...

Page 17: ...nalyzer instruction manual Otherwise proceed to the next step 4 Keeping the donut portion of the sensor immersed in the grab sample clamp the sensor so that it is as far away as possible from the sides and bottom of the beaker yet still in solution You can use a clamp and stand assembly It is important to keep the sensor away from the sides of the beaker because the cur rent and magnetic fields of...

Page 18: ...nly maintenance required during the normal operational life of the sensor is to ensure that there are no significant deposits accumulating on the sen sor 4 2 TROUBLESHOOTING Please refer to Sensor Resistance Check Table 4 1 and Temperature vs Resistance Table 4 2 TABLE 4 1 Sensor Resistance Check Resistance listed are for TC values at 25 C 77 F See Table 4 2 for other temperature values FOR OPTION...

Page 19: ...re C F Resistance Ω 18 64 4 2910 19 66 2 2920 20 68 0 2930 21 69 8 2950 22 71 6 2960 3K 1181T 23 73 4 2970 24 75 2 2990 25 77 0 3000 26 78 8 3010 27 80 6 3030 29 84 2 3050 30 86 0 3060 18 64 4 106 9 19 66 2 107 3 20 68 0 107 7 21 69 8 108 1 22 71 6 108 4 100 Ω RTD 23 73 4 108 9 1054A T 1054B T 2054T 54C 24 75 2 109 2 81T 2081T 3081T and 4081T 25 77 0 109 6 26 78 8 110 0 27 80 6 110 4 ...

Page 20: ...refully package the materials and enclose your Letter of Transmittal see Warranty If possible pack the materials in the same manner as they were received 4 Send the package prepaid to Rosemount Analytical Inc Uniloc Division Uniloc Division 2400 Barranca Parkway Irvine CA 92606 Attn Factory Repair RMA No ____________ Mark the package Returned for Repair Model No ____ 5 3 NON WARRANTY REPAIR The fo...

Page 21: ..._______ ______________________________________________________________________________________________________ ______________________________________________________________________________________________________ WARRANTY REPAIR REQUESTED n n YES REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO ________________________________________ CUSTOMER PURCHASE ORDER NO ___________________________________...

Page 22: ...e the right answers right now THE AMERICAS HEADQUARTERS Emerson Process Management Rosemount Analytical Inc Liquid Center of Excellence 2400 Barranca Parkway Irvine CA 92606 Phone 1 949 757 8500 Toll Free 1 800 854 8257 Fax 1 949 474 7250 ASIA PACIFIC Emerson Process Management Asia Pacific Private Ltd 1 Pandan Crescent Singapore 0512 Republic of Singapore Phone 65 777 8211 Fax 65 777 0947 EUROPE ...

Page 23: ...s or damage directly or indirectly arising from the use of the equipment or goods from breach of any warranty or from any other cause All other warranties expressed or implied are hereby excluded IN CONSIDERATION OF THE STATED PURCHASE PRICE OF THE GOODS SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY NO OTHER WARRANTIES ARE GRANTED INCLUDING BUT NOT LIMITED TO EXPRESS AND IMPLIED WARRANTIES ...

Page 24: ... ON LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http www raihome com Specifications subject to change without notice Emerson Process Management 2400 Barranca Parkway Irvine CA 92606 USA Tel 949 757 8500 Fax 949 474 7250 http www rosemountanalytical com Rosemount Analytical Inc 2012 ...

Reviews: