2
Install
2.1
Installing the sensor
Prerequisites
1. 2 inch Tri-Clamp
2. 2 inch type 1 gasket
3. 2 inch tank ferrule or tee
About this task
The sensor may be installed in either a tank or pipe using a customer-supplied Tri-Clamp
and tee assembly. Keep at least 1 in. (25 cm) between the sensor and the pipe wall. If
clearance is too small, calibrate the sensor in place. Ensure that the sensor is completely
submerged in liquid.
Figure 2-1: Rosemount 225 dimensional drawing
Procedure
1. Mount the sensor in the pipe.
2. Mount the sensor in a vertical pipe run with flow from top to bottom.
If the sensor must be mounted in a horizontal pipe run, orient the sensor in the
3 o'clock or 9 o'clock position.
3. Ensure that the sensor is completely submerged in liquid to the flange.
Reference Manual
Install
00809-0100-3225
August 2018
Rosemount 225 Sensors
11