125
Reference Manual
00809-0100-2240, Rev DA
October 2017
6.3.4
Measurement status for the WLS
shows measurement status for a water level sensor connected to a Rosemount 2240S
Multi-input Temperature Transmitter.
You can view status registers by using the View Diagnostic Registers function or the View Input Registers
function in TankMaster WinSetup, see
Table 6-7. Measurement status for water level sensor
Message
Description
Status
Input register no. 500
(1)
.
Bit 0: No Device Connected
Bit 1: Device reports value under or above 15%
Bit 2: Saturated Low
Bit 3: Saturated High
Bit 4: Relative Pressure
Bit 7: Saturated
Bit 8: In Calibration Mode
Bit 11: Frozen Value
Bit 12: Option Not Available
Bit 13: Power Up
Bit 14: SW HW Error
Bit 15: Invalid
1.
The register number refers to the internal Input Register in the Rosemount 2240S database.
Primary Variable (PV)
Input register no. 502.
Primary value from the connected sensor
Unit
Input register no. 504.
Measurement unit such as Feet, Meters, Inches etc.
7 Bars
8 Millibars
10 Kilograms_per_square_centimeter
11 Pascals
12 Kilopascals
13 Torr
14 Atmospheres
44 Feet
45 Meters
47 Inches
48 Centimeters
49 Millimeters
250 Not_Used
251 None
252 Unknown
253 Special
Summary of Contents for Rosemount 2240S
Page 2: ......
Page 10: ...viii Reference Manual 00809 0100 2240 Rev DA Title Page October 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2240 Rev DA Introduction October 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2240 Rev DA Overview October 2017 Overview ...
Page 192: ...182 Reference Manual 00809 0100 2240 Rev DA Index October 2017 Index ...
Page 193: ......