67
Reference Manual
00809-0100-2230, Rev CA
July 2017
Function check alerts
A Function Check alert indicates that the device is temporary non-valid due to some activities, for
example maintenance, on the device.
There are five parameters associated with FD_CHECK_ALM, they are described below.
FD_CHECK_MAP
The FD_CHECK_MAP parameter contains a list of informative conditions that do not have a direct impact
on the primary functions of the device. Below is a list of the conditions:
1. Check function
FD_CHECK_MASK
The FD_CHECK_MASK parameter will mask any of the failed conditions listed in FD_CHECK_MAP. A bit
on means the condition is masked out from alarming and being broadcast to the host through the alarm
parameter.
FD_CHECK_PRI
FD_CHECK_PRI designates the alarming priority of the FD_CHECK_ALM, see
. The default is 0 and the recommended values are 1 or 2.
FD_CHECK_ACTIVE
The FD_CHECK_ACTIVE parameter displays which of the conditions is active.
FD_CHECK_ALM
FD_CHECK_ALM is an alarm indicating that the device output is temporary invalid due to on-going work
on the device.
4.12.4
Recommended actions for alerts
The FD_RECOMMEN_ACT and RECOMMENDED_ACTION parameters display text strings that will give a
recommended course of action to take based on which type and which specific event of the alerts that is
active (See
).
Summary of Contents for Rosemount 2230
Page 1: ...Reference Manual 00809 0100 2230 Rev CA July 2017 Rosemount 2230 Graphical Field Display ...
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 2230 Rev CA Title Page July 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2230 Rev CA Introduction July 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2230 Rev CA Overview July 2017 Overview ...
Page 44: ...34 Reference Manual 00809 0100 2230 Rev CA Installation July 2017 Installation ...
Page 154: ...144 Reference Manual 00809 0100 2230 Rev CA Index July 2017 Index ...
Page 155: ......