5
Troubleshooting
5.1
Troubleshooting
Figure 5-1: Disconnect Wires From Transmitter Before Measuring Resistance
Table 5-1: Resistance vs. Temperature for Temperature Compensation (PT-100 RTD)
Temperature
Resistance
10 °C (50 °F)
103.9 Ω
20 °C (68 °F)
107.8 Ω
25 °C (77 °F)
109.7 Ω
30 °C (86 °F)
111.7 Ω
40 °C (104 °F)
115.5 Ω
50 °C (122 °F)
119.4 Ω
5.1.1
Checking the toroid operation
Use the following procedure to check the operation of the toroids.
Procedure
1. Disconnect the shorting strap from one of the outer flanges.
The conductivity reading will drop to zero.
2. Pass a short piece of heavy gauge wire through the space between the toroid
assembly and the pipe.
3. Connect the ends of the wire to a resistance decade box.
4. Turn off temperature correction in the transmitter.
Reference Manual
Troubleshooting
00809-0100-3222
September 2018
Rosemount 222 Sensors
21
Summary of Contents for Rosemount 222
Page 2: ...2 ...
Page 4: ...4 ...
Page 6: ...Contents Reference Manual September 2018 00809 0100 3222 6 Reference Manual ...
Page 20: ...Calibration Reference Manual September 2018 00809 0100 3222 20 Reference Manual ...
Page 24: ...Accessories Reference Manual September 2018 00809 0100 3222 24 Reference Manual ...