54
Reference Manual
00809-0400-4101, Rev BA
Operation and Maintenance
January 2018
Operation and Maintenance
AMS Device Manager
1. Right click the device and, under the
Method
drop down menu, move cursor over
Calibrate
and select
Restore Factory Calibration
.
2. Select
Next
to set the control loop to manual.
3. Select
Analog Output Trim
under
Select trim to recall
and click
Next
.
4. Follow screen prompts to recall analog output trim.
LOI
for LOI instructions.
Figure 5-7. Recall Factory Trim – Analog Output with LOI
5.7
Switching HART revision
Some systems are not capable of communicating with HART
®
Revision 7 devices. The following
procedures list how to change HART revisions between HART Revision 7 and HART Revision 5.
5.7.1
Generic menu
If the HART configuration tool is not capable of communicating with a HART Revision 7 device, it should
load a generic menu with limited capability. The following procedures allow for switching between
HART Revision 7 and HART Revision 5 from a generic menu.
1. Locate “Message” field.
a. To change to HART Revision 5, Enter:
HART5
in the message field.
b. To change to HART Revision 7, Enter:
HART7
in the message field.
5.7.2
Field Communicator
From the HOME screen, enter the Fast Key sequence and follow steps within the Field Communicator to
complete the HART revision change.
From the
HOME
screen, enter the Fast Key sequence
HART5
HART7
Device Dashboard Fast Keys
2, 2, 5, 2, 4
2, 2, 5, 2, 3
EXTENDED MENU
CALIBRAT
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN PV
TAG
ALARM SAT VALUES
PASSWORD
SIMULATE
HART REV
BACK TO MENU
EXIT MENU
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXTENDED MENU
EXIT MENU
CALIBRAT
ZERO TRIM
LOWER TRIM
UPPER TRIM
ANALOG TRIM
FACTORY RECALL
FACTORY RECALL
BACK TO MENU
EXIT MENU
FACTORY RECALL
SENSOR RECALL
ANALOG RECALL
ANALOG RECALL
BACK TO MENU
EXIT MENU
Summary of Contents for Rosemount 2051G
Page 2: ......
Page 10: ...2 Reference Manual 00809 0400 4101 Rev BA Introduction January 2018 Introduction ...
Page 34: ...26 Reference Manual 00809 0300 4410 Rev BA Configuration January 2018 Configuration ...
Page 72: ...64 Reference Manual 00809 0400 4101 Rev BA Troubleshooting January 2018 Troubleshooting ...
Page 89: ......