9
Reference Manual
00809-0300-4410, Rev BA
Configuration
January 2018
Configuration
2.5
Configuration overview
This section contains information on commissioning and tasks that should be performed on the bench
prior to installation, as well as tasks performed after installation as described in
Field Communicator, AMS
Device Manager, and Local Operator Interface (LOI) instructions are given to
perform configuration functions. For convenience, Field Communicator Fast Key sequences are labeled
“Fast Keys,” and abbreviated LOI menus are provided for each function below.
Full Field Communicator menu trees and Fast Key sequences are available in
Communicator Menu Trees and Fast Keys
. LOI menu trees are available in
2.6
Configuration basics
The Rosemount 2051G can be configured either before or after installation. Configuring the transmitter
on the bench using either a Field Communicator, AMS Device Manager, or LOI ensures all transmitter
components are in working order prior to installation. Verify that the security switch is set in the unlock
position (
) in order to proceed with configuration. See
for switch location.
2.6.1
Configuring on the bench
To configure on the bench, required equipment includes a power supply, and a Field Communicator,
AMS Device Manager, or an LOI (option M4). Wire equipment as shown in
below. To ensure
successful HART communication, a resistance of at least 250
Ω
s must be present between the
transmitter and the power supply, see
for details. Connect the Field
Communicator leads to the terminals labeled “COMM” on the terminal block.
Figure 2-4. Wiring the Transmitter (4–20 mA HART)
A. Vdc supply
B. R
L
≥
250 (necessary for HART communication only)
Set all transmitter hardware adjustments during commissioning to avoid exposing the transmitter
electronics to the plant environment after installation.
A
B
Summary of Contents for Rosemount 2051G
Page 2: ......
Page 10: ...2 Reference Manual 00809 0400 4101 Rev BA Introduction January 2018 Introduction ...
Page 34: ...26 Reference Manual 00809 0300 4410 Rev BA Configuration January 2018 Configuration ...
Page 72: ...64 Reference Manual 00809 0400 4101 Rev BA Troubleshooting January 2018 Troubleshooting ...