Table 3-2: RFW Upper Housing Torque Values
Materials (nuts and
botls)
Bolt thread size
Class
Torque
SST
3/8–24 NF
150/300/600 lb
23 ft-lb
CS
3/8–24 NF
150/300/600 lb
CS
½
–20 NF
900/1500 lb
(1)
53 ft-lb for 4.1-in. diaphragm size.
Note
This is the specification for connecting the remote seal to the lower housing, not the
torque specification for the lower housing onto the process flanged connection. The lower
housing bolts should be torqued to the applicable flange requirements.
3.6.1
4.1-in. (104 mm) diaphragm diameter option
The standard diaphragm size for the RFW remote flanged seal is 2.4-in. A larger, 4.1-in.
(104 mm), diaphragm size is offered for small spans to reduce temperature error when
taking process measurements.
3.6.2
Parts required for installation
The parts required to install the RFW remote flanged seal will be defined here. Prior to
installation, you will need the following:
•
Torque wrench
•
Mounting hardware (end-user-supplied)
•
Gasket (lower housing to process flange) (end-user-supplied)
•
Gasket (remote seal to lower housing) (Emerson-supplied)
Note
Typically, the stud bolt design includes Emerson-supplied stud bolts.
Verify the gasket material is appropriate for the application.
Inspect bolts to ensure the material is compatible with industry standards per the
application such as ASME PCC-1.
3.6.3
Standard design installation steps
It is recommended that two people install the standard design RFW remote flanged seal to
ensure proper alignment during installation.
Follow these steps to install the standard design RFW remote flanged seal on an existing
process flange:
Procedure
1. Remove the protective cover from the diaphragm of the remote seal. Use extreme
caution during installation to ensure the diaphragm is not damaged.
Installation
Reference Manual
May 2020
00809-0100-4002
32
Rosemount DP Level
Summary of Contents for Rosemount 1199
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual May 2020 00809 0100 4002 8 Rosemount DP Level ...
Page 22: ...Remote Seal Systems Reference Manual May 2020 00809 0100 4002 22 Rosemount DP Level ...
Page 74: ...Installation Reference Manual May 2020 00809 0100 4002 74 Rosemount DP Level ...
Page 92: ...Fill Fluids Reference Manual May 2020 00809 0100 4002 92 Rosemount DP Level ...