IMPORTANT NOTICE – ERRATA
Model 1153 Series H Product Manual 00809-0100-4631 Rev BA (April 2007)
No.
Affected
Pages
Description of Change
Effect.
Date
1
6-6
Process Flagel –
CF3M
(Cast version of
316L
SST)
Drain/Vent Valves –
316L
SST
Process Connections – 3/8-inch Swagelok compression fitting,
316L SST
(1/4-18 NPT
optional)
10/21/09
2
3-6
Change the first paragraph on the page to read as follows:
“Damping electronics are available as an option. Transmitters with standard electronics
can be retrofitted with the adjustable damping feature by changing out both the amplifier
board and the calibration board. Please reference Table 6-2 (Parts List Rosemount
1154DH, 1154HH, and 1154SH) for the applicable part numbers.”
4/13/12
3
6-1
Steam Pressure/Temperature specification is updated as follows:
Steam Pressure/Temperature
Accuracy as specified below, during and after sequential exposure to steam at the
following temperatures and pressures, concurrent with chemical spray for the first 24
hours.
420 °F (215.6 °C), 85 psig for 3 minutes
350 °F (176.6 °C), 85 psig for 7 minutes
320 °F (160 °C), 75 psig for 8 hours
265 °F (129.4 °C), 24 psig for 56 hours
Range codes 4-8: Accuracy within ±(1.0% of upper range limit + 1.0% of span)
(1)
Range code 9: Accuracy within ±(2.0% of upper range limit + 0.5% of span)
Note
(1) For Range codes 4-8, if pressure applied at the 4mA point is greater than 50% of the upper range limit, the
specification is ±2% of upper range limit.
4/13/12
4
6-8
Table 6-2, “Parts List Rosemount 1154DH, 1154HH, and 1154SH”, is updated as
follows:
Amplifier Circuit Board, Output Code R:
In all locations, replaced “01154-0001-0005” with “01154-0153-0001”
Amplifier Circuit Board with Damping, Output Code R:
In all locations, replaced “01154-0021-0004” with “01154-0156-0001”
Amplifier Circuit Board, Output Code R, N0026:
In all locations, replaced “01154-0001-0006” with “01154-0153-0002”
4/13/12
5
6-11
Table 6-2, “Parts List Rosemount 1154DH, 1154HH, and 1154SH”, in all locations the
following part numbers are updated:
Amplifier Circuit Board, Output Code R:
“01154-0153-0001” is replaced by “01154-0263-0001”
Amplifier Circuit Board with Damping, Output Code R:
“01154-0156-0001” is replaced by “01154-0266-0001”
Amplifier Circuit Board, Output Code R, N0026:
“01154-0153-0002” is replaced by “01154-0263-0002”
6/9/17
Summary of Contents for Rosemount 1154DH
Page 6: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 1 2 ...
Page 10: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H TOC 4 ...
Page 12: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 1 2 ...
Page 52: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 6 10 ...
Page 55: ......