Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13
Model MicroCEM
Calibration Setting
Description
O2ZeroDriftLimit
O2 Allowed Zero Drift Limit.
COZeroDriftLimit
CO Allowed Zero Drift Limit.
NOXZeroDriftLimit
NOx Allowed Zero Drift Limit.
OSMidDriftLimit
O2 Allowed Mid Drift Limit.
COMidDriftLimit
CO Allowed Mid Drift Limit.
NOXMidDriftLimit
NOx Allowed Mid Drift Limit.
O2SpanDriftLimit
O2 Allowed Span Drift Limit.
COSpanDriftLimit
CO Allowed Span Drift Limit.
NOXSpanDriftLimit
NOx Allowed Span Drift Limit.
If the drift exceeds the allowed amount a drift alarm
will occur, and the readings on the channel will no
longer be valid until a successful calibration is com-
pleted.
O2Slope
Default value for the O2 slope
O2Offset
Default value for the O2 offset
COSlope
Default value for the CO slope
COOffset
Default value for the CO offset
NOXSlope
Default value for the NOx slope
NOXOffset
Default value for the NOx offset
Initial values for the emissions conversion slope and
offset used on a new system before the first Cali-
bration is performed. These values should be set
manually before the first auto-calibration is per-
formed.
O2SpanDef
Default O2 Span Calibration Gas value
O2MidSpanDef
Default O2 Mid Span Calibration Gas value
COSpanDef
Default CO Span Calibration Gas value
COMidSpanDef
Default CO Mid Span Calibration Gas value
NOXSpanDef
Default NOx Span Calibration Gas value
NOXMidSpanDef
Default NOx Mid Span Calibration Gas value
Gas1Allocation
Default Gas 1 allocation
Gas2Allocation
Default Gas 2 allocation
Gas3Allocation
Default Gas 3 allocation
Default allocation for O2/NOx/CO. 1 = Zero, 2 = Mid
Span, 3 = Span. Example: 133 =
O2=Zero,NOx=Span,CO=Span
AutoCalFrequency
Default Auto-Calibration frequency in hours and minutes (example: 24:00).
AutoCalTime
Default Auto-Calibration time in hours and minutes (military time).
Purge1
Default auto-calibration Purge 1 value in seconds.
Gas1
Default auto-calibration gas 1 time in seconds.
Gas2
Default auto-calibration gas 2 time in seconds.
Gas3
Default auto-calibration gas 3 time in seconds.
Purge2
Default auto-calibration Purge 2 value in seconds.
Table 3-3. Factory Settings – Calibration