Installation Manual
25
Installation
3.10
Grounding
The sensor/core processor assembly (see
) or the sensor alone (see
) must be grounded. To ground these components, see the sensor
installation manual.
If your installation includes a remote core processor (see
and
), it must
be grounded. To ground the remote core processor:
•
The core processor has two internal ground screws: one 4-wire ground screw and one
9-wire ground screw. Do not use the 4-wire ground screw. The 9-wire ground screw
may be used (see
). To access the 9-wire ground screw, remove the core
).
•
Use copper wire, 14 AWG (2,0 mm
2
) or larger, for grounding.
•
Keep all ground leads as short as possible, less than 1 ohm impedance.
•
Connect ground leads directly to earth, or follow plant standards.
If your installation includes the MVD Direct Connect I.S. barrier (see
), the barrier is not grounded. Do not ground the barrier.
Figure 3-13 Core processor 9-wire ground screw
Note
This step is required for all MVD Direct Connect installations.
CAUTION!
Improper grounding could cause measurement error. To reduce the risk of measurement
error:
• Ground the meter to earth, or follow ground network requirements for the facility.
• For installation in an area that requires intrinsic safety, refer to the appropriate Micro
Motion approval documentation.
• For hazardous area installations in Europe, refer to standard EN 60079-14 if national
standards do not apply.
9-wire ground screw
Summary of Contents for Micro Motion MVD Direct Connect
Page 1: ...Installation Manual 20002273 Rev AA July 2014 Micro Motion MVD Direct Connect Meters ...
Page 4: ...4 Direct Connect Meters ...
Page 6: ...6 Direct Connect Meters Before you begin ...
Page 26: ...26 Direct Connect Meters Installation ...
Page 29: ...Installation Manual 29 MVD Direct Connect communications ...