Appendix A
Default values and ranges
Topics covered in this appendix:
A.1
Default values and ranges
See Table A-1 for the default values and ranges for the most frequently used transmitter settings.
Table A-1
Transmitter default values and ranges
Type
Setting
Default
Range
Comments
Flow direction
Forward
Flow damping
0.8 sec
0.0 – 40.96
sec
User-entered value is corrected
to nearest lower value in list
of preset values. For gas
applications, Micro Motion
recommends a minimum value
of 2.56.
Mass flow units
g/s
Mass flow cutoff
0.0 g/s
Recommended setting is 5%
of the sensor’s rated maximum
flowrate.
Volume flow type
Liquid
Volume flow units
L/s
Flow
Volume flow cutoff
0/0 L/s
0.0 – x L/s
x is obtained by multiplying the
flow calibration factor by 0.2,
using units of L/s.
Mass factor
1
Density factor
1
Meter factors
Volume factor
1
Configuration and Use Manual
203