Before performing the calibration, record your current calibration parameters. If the
calibration fails, restore the known values.
Restriction
For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be
performed on water.
Procedure
See
D1 and D2 density calibration using the Field Communicator
Figure 11-9:
Enter density of D1 fluid
Density Calibration
Complete
message
Calibration in Progress
message
D1 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D1 fluid
Fill sensor with D2 fluid
D2 calibration
OK
Dens Pt 1
Calibration method
executes
OK
Home
Enter density of D2 fluid
Density Calibration
Complete
message
Calibration in Progress
message
OK
Dens Pt 2
Calibration method
executes
OK
Home
Done
On-Line Menu >
Service Tools >
Maintenance >
Density Calibration
Service Tools >
Maintenance >
Density Calibration
Postrequisites
If you disabled
LD Optimization
before the calibration procedure, re-enable it.
Measurement support
198
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......