D3 or D3 and D4 density calibration using ProLink III
Figure 11-1:
D3 Calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration - Point 3
Device Tools >
Calibration >
Density Calibration >
Density Calibration - Point 4
Enter density of D3 fluid
Start Calibration
Start Calibration
Enter density of D4 fluid
Close
Close
Done
D4 Calibration
Fill sensor with D4 fluid
11.9.2
Perform a D3 or D3 and D4 density calibration using the
Field Communicator
See the following flowchart.
Measurement support
Configuration and Use Manual
157