Some alarms can be reclassified. For example:
•
The default severity level for Alarm A20 (calibration factors unentered) is
Fault
, but you can
reconfigure it to either
Informational
or
Ignore
.
•
The default severity level for Alarm A102 (drive over-range) is
Informational
, but you can
reconfigure it to either
Ignore
or
Fault
.
For a list of all status alarms and default severity levels, see Table 8-7. (For more information on
status alarms, including possible causes and troubleshooting suggestions, see Table 11-4.)
To configure alarm severity:
•
Using ProLink II, see Figure C-3.
•
Using the Communicator, see Figure C-5.
Note: This functionality is not available via the display menus.
Table 8-7
Status alarms and severity levels
Alarm code
Communicator message
Default
severity
Configurable
Affected by
fault timeout
ProLink II message
A001
EEprom Checksum Error (Core Processor)
Fault
No
No
(E)EPROM Checksum Error (CP)
A002
RAM Test Error (Core Processor)
Fault
No
No
RAM Error (CP)
A003
Sensor Not Responding (No Tube Interrupt)
Fault
Yes
Yes
Sensor Failure
A004
Temperature sensor out of range
Fault
No
Yes
Temperature Sensor Failure
A005
Input Over-Range
Fault
Yes
Yes
Input Overrange
A006
Transmitter Not Characterized
Fault
Yes
No
Not Configured
A008
Density Outside Limits
Fault
Yes
Yes
Density Overrange
A009
Transmitter Initializing/Warming Up
Fault
Yes
No
Transmitter Initializing/Warming Up
A010
Calibration Failure
Fault
No
No
Calibration Failure
A011
Excess Calibration Correction, Zero too Low
Fault
Yes
No
Zero Too Low
A012
Excess Calibration Correction, Zero too High
Fault
Yes
No
Zero Too High
A013
Process too Noisy to Perform Auto Zero
Fault
Yes
No
Zero Too Noisy
A014
Transmitter Failed
Fault
No
No
Transmitter Failed
Summary of Contents for Micro-Motion 2400S
Page 8: ...vi Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 20: ...12 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 30: ...22 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 42: ...34 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 66: ...58 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 76: ...68 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 138: ...130 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 190: ...182 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 197: ......