background image

Model Solu Comp

®

Xmt-P-FF/FI

F

OUNDATION

®

Fieldbus Two-Wire

pH/ORP Transmitter

Instruction Sheet

PN 51A-Xmt-P-FF-FI/rev.A
January 2006

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!

Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your
particular application. These instruments have been designed, and tested to meet many national and interna-
tional standards. Experience indicates that its performance is directly related to the quality of the installation
and knowledge of the user in operating and maintaining the instrument. To ensure their continued operation to
the design specifications, personnel should read this manual thoroughly before proceeding with installation,
commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not speci-
fied by the manufacturer, the protection provided by it against hazards may be impaired.

• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life;

personal injury; property damage; damage to this instrument; and warranty  invalidation.

• Ensure that you have received the correct model and options from your purchase order. Verify that this man-

ual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.

• For clarification of instructions, contact your Rosemount representative.

• Follow all warnings, cautions, and instructions marked on and supplied with the product.

• Use only qualified personnel to install, operate, update, program and maintain the product. 

• Educate your personnel in the proper installation, operation, and maintenance of the product.

• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national

codes. Only connect the product to electrical and pressure sources specified in this manual.

• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and

procedures can affect the performance and cause unsafe operation of your process.

• All equipment doors must be closed and protective covers must be in place unless qualified personnel are

performing maintenance.

• If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against

hazards may be impaired.

For additional information, please refer to the Instruction Manuals CD shipped with this
product, or visit our website at 

www.emersonprocess.com/raihome/liquid/

.

Summary of Contents for FOUNDATION Solu Comp Xmt-P-FF

Page 1: ... have received the correct model and options from your purchase order Verify that this man ual covers your model and options If not call 1 800 854 8257 or 949 757 8500 to request correct manual For clarification of instructions contact your Rosemount representative Follow all warnings cautions and instructions marked on and supplied with the product Use only qualified personnel to install operate ...

Page 2: ...s a stabilization check on the sensor in each buffer A manual two point calibration is made by immersing the sensor in two different buffer solutions and entering the pH values The microprocessor automatically calculates the slope which is used for self diagnostics An error message will be displayed if the pH sensor is faulty This slope can be read on the display and or manually adjusted if desire...

Page 3: ...y 381 pHE 31 41 52 Glass only 381 Glass and Reference 385 Glass and Reference 389 02 54 389VP 54 Glass only 396 54 62 396VP Glass only 396P 55 396PVP 55 Glass and Reference 396R 396RVP 54 Glass and Reference 397 54 62 Glass only 398 54 62 398VP 54 Glass only 398R 54 62 398RVP 54 Glass only 399 09 62 399VP 09 399 14 Glass only Hx338 Glass only Hx348 Glass only TF396 none ...

Page 4: ...Xm mt tr r Sensor JBox Temperature in C C F 1 Refer to page 5 for installation instructions 2 Wire pH or ORP sensor to the transmitter See Figure 6 for panel mount Figure 7 for pipe or surface mount Refer to the sensor instruction sheet for details 3 Once connections are secure and verified apply power to the transmitter 4 When the transmitter is powered up for the first time Quick Start screens a...

Page 5: ...F FI has four knock outs The pipe surface mount transmitter has two knockouts One conduit opening is for the power out put cable the other opening is for the sensor cable Figure 1 shows how to remove a knockout The knockout grooves are on the outside of the case Place the screwdriver blade on the inside of the case and align it approximately along the groove Rap the screwdriver sharply with a hamm...

Page 6: ...MODEL Xmt P INSTALLATION 6 FIGURE 3 Panel Mount Installation Access to the wiring terminals is through the rear cover Four screws hold the cover in place Panel Mounting MILLIMETER INCH ...

Page 7: ...MODEL Xmt P INSTALLATION 7 FIGURE 4 Pipe Mount Installation The front panel is hinged at the bottom The panel swings down for access to the wiring terminals Pipe Mounting MILLIMETER INCH ...

Page 8: ...8 MODEL Xmt P INSTALLATION FIGURE 5 Surface Mount Installation The front panel is hinged at the bottom The panel swings down for access to the wiring terminals Surface Mounting MILLIMETER INCH ...

Page 9: ...9 MODEL Xmt P INSTALLATION FIGURE 6 Loop Power and Sensor Wiring Panel Mount ...

Page 10: ...10 FIGURE 7 Loop Power and Sensor Wiring Pipe Surface Mount ...

Page 11: ...MAIN MENU MENU TREE FOR MODEL SOLU COMP Xmt P FF FI TRANSMITTER 11 ...

Page 12: ...N LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http www raihome com Specifications subject to change without notice Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine CA 92606 USA Tel 949 757 8500 Fax 949 474 7250 http www raihome com Rosemount Analytical Inc 2006 ...

Reviews: