DVC6000 Digital Valve Controllers
September 2013
5-4
Manual Calibration (1-3-1-2)
It is recommended that you adjust the relay before
manually calibrating travel. For additional information
refer to Relay Adjustment in this section.
Note
Relay Adjustment is only available for
the double-acting relay (Relay A).
Two procedures are available to manually calibrate
travel:
Analog Adjust
Digital Adjust
Analog Calibration Adjust
Select Calibrate, Travel Calibration, Manual
Calibration and Analog Adjust. Connect a variable
current source to the instrument LOOP + and LOOP
−
terminals. The current source should be capable of
generating 4 to 20 mA. Follow the prompts on the
Field Communicator display to calibrate the
instrument’s travel in percent.
Note
0% Travel = Valve Closed
100% Travel = Valve Open
1. Adjust the input current until the valve is near
mid-travel. Press OK.
2. If the feedback connection is Rotary - All, SStem -
Roller, go to step 6. If the feedback connection is
SStem - Standard, you are prompted to set the
crossover point. Adjust the current source until the
feedback arm is 90
°
to the actuator stem, as shown in
Table 5-1. Auto Calibrate Travel Error Messages
Error Message
Possible Problem and Remedy
Input current must
exceed 3.8 mA for
calibration.
The analog input signal to the instrument
must be greater than 3.8 mA. Adjust the
current output from the control system or the
current source to provide at least 4.0 mA.
Place Out Of Service
and ensure Calibrate
Protection is disabled
before calib.
The Instrument Mode must be Out of Service
and the Protection must be None before the
instrument can be calibrated. For information
on changing instrument protection and mode,
see the beginning of this section.
Calibration Aborted.
An end point was not
reached.
The problem may be one or the other of the
following:
1. The tuning set selected is too low and the
valve does not reach an end point in the
allotted time. Press the Hot Key, select
Stabilize/Optimize then Increase Response
(selects next higher tuning set).
2. The tuning set selected is too high, valve
operation is unstable and does not stay at an
end point for the allotted time. Press the Hot
Key, select Stabilize/Optimize then Decrease
Response (selects next lower tuning set).
Invalid travel value.
Check travel sensor
and feedback arm
adjustments, and inst
supply press. Then,
repeat Auto Calib.
Prior to receiving this message, did the
instrument output go from zero to full supply?
If not, verify instrument supply pressure by
referring to the specifications in the
appropriate actuator instruction manual. If
supply pressure is correct, check instrument
pneumatic components (I/P converter and
relay).
If the instrument output did go from zero to
full supply prior to receiving this message,
then verify proper mounting by referring to the
appropriate mounting procedure in the
Installation section.
Verify travel sensor adjustment by performing
the appropriate Travel Sensor Adjust
procedure in the Calibration section.
Making the crossover adjustment with the
valve positioned at either end of its travel will
also cause this message to appear.
Aborting due to
response code or
device status.
The instrument may have been taken out of
service by a primary master. Put the
instrument into service with the primary
master or cycle the power off and on.
Note
In steps 3 through 7, the accuracy of
the current source adjustment affects
the position accuracy.
3. Adjust the current source until the valve is at 0%
travel, then press OK.
4. Adjust the current source until the valve is at 100%
travel, then press OK.
5. Adjust the current source until the valve is at 50%
travel, then press OK.
5
Summary of Contents for Fisher FIELDVUE DVC6000
Page 2: ......
Page 22: ...DVC6000 Digital Valve Controllers September 2013 1 12 1 ...
Page 82: ...DVC6000 Digital Valve Controllers September 2013 4 22 4 ...
Page 94: ...DVC6000 Digital Valve Controllers September 2013 5 12 5 ...
Page 139: ...Loop Schematics Nameplates September 2013 B 1 B B Appendix B Loop Schematics Nameplates B ...
Page 161: ......