5-7
L6118EA
14. Sub-Brightness Adjustment
Purpose:
To get proper brightness.
Symptom of Misadjustment:
If Sub-Brightness is
incorrect, proper brightness cannot be obtained by
adjusting the Brightness Control.
Note:
IQW Setup level --- 7.5 IRE
Use Service remote control unit
1. Enter the Service mode. (See page 5-1)
Then input IQW signal from RF Input.
2. Press "MENU" button on the Service remote control
unit and select "BRT" mode. (Display changes
"BRT," "CNT," "CLR," and "TNT" cyclically when
MENU button is pressed.) Press "CH
o
/
p
" buttons
so that the bar is just visible (See above figure.)
3. Turn the power off and on again, using the main
power button on the TV unit.
15. Focus Adjustment
Purpose:
Set the optimum Focus.
Symptom of Misadjustment:
If Focus Adjustment is
incorrect, blurred images are shown on the display.
Note:
Focus VR (FBT) - Main CBA FBT=Fly Back Trans-
former
1. Operate the unit more than 30 minutes
2. Face the unit to the East and degauss the CRT
using a degaussing coil.
3. Input the Monoscope Pattern.
4. Adjust the Focus Control on the FBT to obtain a
clear picture.
The following 2 adjustments normally are not
attempted in the field. They should be done only
when replacing the CRT then adjust as a prepara-
tion.
16. Purity Adjustment
Purpose:
To obtain pure color.
Symptom of Misadjustment:
If Color Purity Adjust-
ment is incorrect, large areas of color may not be
properly displayed.
1. Set the unit facing the East.
2. Operate the unit for over 30 minutes before adjust-
ing.
3. Fully degauss the unit using an external degauss-
ing coil.
4. Loosen the screw on the Deflection Yoke Clamper
and pull the Deflection Yoke back away from the
screen. (See Fig. 6)
5. Loosen the Ring Lock and adjust the Purity Mag-
nets so that a red field is obtained at the center of
the screen. Tighten Ring Lock. (See Fig. 5,6)
6. Slowly push the Deflection Yoke toward the bell of
the CRT and set it where a uniform red field is ob-
tained.
7. Tighten the clamp screw on the Deflection Yoke.
Test Point
Adj. Point
Mode
Input
---
CH
o
/
p
buttons
RF
IQW
Tape
M. EQ.
Spec.
---
Pattern
Generator
See below
Figure
Test Point
Adj. Point
Mode
Input
---
Focus Control
RF
Mono-
scope
Tape
M. EQ.
Spec.
---
Pattern
Generator
See below
White
Black
Fig. 4
This bar
just
visible
Test Point
Adj. Point
Mode
Input
---
Deflection
Yoke Purity
Magnet
---
Red
Color
Tape
M. EQ.
Spec.
---
Pattern
Generator
See below.
Figure
GREEN
RED
BLUE
Fig. 5