Ultrasonic Power Problems
Table 8.16 Troubleshooting Ultrasonic Power Problems
Problem
Check
Solution
Ultrasonic power delivered
to horn; no indication on
bar graph.
Check connector cables,
replace if failed.
Replace defective
cables.
Test power supply.
See
No ultrasonic power
generated when Test key
pressed; no Alarm indicator.
Failed or missing stack.
Replace.
RF cable unplugged or failed;
replace if failed.
Plug in or replace.
Test power supply (see
).
If defective, send unit
for repair.
Unable to adjust amplitude
using the front panel
keypad.
Register setting configured to
“External Amplitude Control”
Reset if required, See
.
Unable to remote control.
User I/O cable
Repair or replace.
Customer’s switching device
Test/inspect/repair/
replace.
Summary of Contents for DCX S
Page 4: ...iv 100 412 183 REV 13 ...
Page 22: ...12 100 412 183 REV 13 ...
Page 36: ...26 100 412 183 REV 13 ...
Page 47: ...100 412 183 REV 13 37 4 3 Declaration of Conformity Figure 4 1 Declaration of Conformity ...
Page 50: ...40 100 412 183 REV 13 ...
Page 54: ...44 100 412 183 REV 13 Figure 5 1 DCX S Power Supply Benchtop Dimensional Drawing ...
Page 86: ...76 100 412 183 REV 13 ...
Page 87: ...100 412 183 REV 13 77 Chapter 6 Converters and Boosters 6 1 Converters and Boosters 78 ...
Page 120: ...110 100 412 183 REV 13 ...
Page 137: ...100 412 183 REV 13 127 8 5 Circuit Diagram Figure 8 2 Interconnect Diagram Power Supply ...
Page 144: ...134 100 412 183 REV 13 ...
Page 145: ...100 412 183 REV 13 135 Appendix A Timing Diagrams A 1 Timing Diagrams 136 ...
Page 147: ...100 412 183 REV 13 137 Appendix B Signal Diagrams B 1 Signal Diagrams 138 ...
Page 154: ...144 100 412 183 REV 13 ...