installed but with out the bolts installed, may cause the electronics to fall and cause
personal injury or electronics damage.
Figure 2-11: Incorrect sling attachment
7. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock
loading ever occurs, the slings must be inspected per manufacturer's procedures
prior to being placed in any further service.
2.5
Mounting requirements in heated or cooled
pipelines
The ambient operating temperature of the Daniel 3814 Liquid Ultrasonic Flow Meter
electronics (i.e. Flameproof enclosure and Intrinsically safe base enclosure) is -40 °C (-40
°F) to +60 °C (+140 °F). If the meter is installed into a pipeline which is heated or cooled
outside this temperature range it is necessary to remove the electronics housing from the
meter body (i.e. Spool piece acting as process fluid conduit) and mount it next to the
meter body on a pipe stand or other rigid structure.
Extended length transducer cables (P/N 1-360-01-601, 15 ft. long) shall be used to
connect the Daniel 3814 Liquid Ultrasonic Flow Meter electronics to the transducers
Mechanical Installation
Installation manual
May 2019
3-9000-760
46
3814 Liquid Ultrasonic Flow Meter