Instruction Manual
D103621X012
4320
January 2018
8
Table 4. Device Variables
INDEX
DEVICE VARIABLE
DESCRIPTION
DD
Published To Gateway
0
Position
PERCENT_OF_SPAN
Travel
(% of Calibration Output Span)
1
Closed/Low Limit Switch
CLOSED_TRIGGER
Reports Closed/Low condition
(1 = Tripped, 0 = Not Tripped)
2
Opened/High Limit Switch
OPEN_TRIGGER
Reports Opened/High condition
(1 = Tripped, 0 = Not Tripped)
3
Temperature
DEVICE_TEMPERATURE
Electronics Module Temperature
(Degrees C)
4
Supply Voltage
BATTERY_VOLTAGE
Power Supply Output
(Volts DC)
5
Set Point (on/off control only)
SET_POINT
Set Point
(1 = Closed, 2 = Open)
6
Switch States
SWITCH_STATE
Combines information in variables 1
and 2
(0 = Part Open, 1 = Closed,
2 = Opened, 3 = Unknown,
4 = Mid, 5 = High, 6 = Low)
7
Last Close Stroke Time
CLOSE_STROKE_TIME
Time the on/off controller took to previously close the valve
8
Last Open Stroke Time
OPEN_STROKE_TIME
Time the on/off controller took to previously open the valve
9
Timestamp 0
MESSAGE_0_TIMESTAMP
Timestamp of burst message 0
10
Timestamp 1
MESSAGE_1_TIMESTAMP
Timestamp of burst message 1
11
Timestamp 2
MESSAGE_2_TIMESTAMP
Timestamp of burst message 2
12
Opened/High Dwell
OPEN_DWELL_TIME
The time the state has been/previously was in the open state.
13
Closed/Low Dwell
CLOSED_DWELL_TIME
The time the state has been/previously was in the closed state.
14
Transition Dwell
PARTIALLY_OPEN_DWELL_TIME
The time the state has been/previously was in the partially open state.
15
Cycle Counter
CYCLE_COUNTS
Cycle counted by the device
16
Scaled Position
16
Variable that can be configured to present the normalized (%) Position
signal in engineering units.
243
Battery Life Remaining
BATTERY_LIFE
Estimated battery life remaining
(days) - meaningless with
External Power option
244
PV % Range
PERCENT_RANGE
Copy of Variable 0
(%)
245
Loop Current
CURRENT
Always Indeterminate (not a number)
(mA)
INDEX
DISCRETE VARIABLES
DESCRIPTION
DD
Published To Gateway
0
Discrete Switch State
DISCRETE_SWITCH_STATE
Current state of the switch variable
(6 = Closed, 17 = Opening,
18 = Closing, 46 = Opened
48 = Unknown, 52 = Part-Open,
53 = Mid, 54 = High, 55 = Low)
1
Discrete Set Point
(On/Off Control only)
DISCRETE_SET_POINT
Target value for the controlled variable
(4 = Close, 5 = Open,
6 = Closed, 46 = Opened)
1. 1410 gateway version 4.4.28 or later or 1420 gateway version 4.4.15 or later is required.