12
“PROPORTIONAL” mode (e.g. conductivity)
This working mode ca be set for OUT1 and OUT2 outputs.
Proportional mode set the controller to operate using a calculated percentage based on a 1000 seconds scale between
two set values that enable or disable the Conductivity pump. To use this mode move cursor on “Working Mode”. Press
the wheel and select it.
PROPORTIONAL MODE between 10mS (0%) and 25mS (100%). [100 seconds scale]
This mode operates Conductivity pump for a value lower that 10mS with maximum dosing capacity and it will stop
Conductivity pump for a reading value higher than 25mS. Between these two vales (17,5mS) the pump will be
controlled into proportional mode (50seconds ON, 50seconds OFF).