background image

1

OPERATING INSTRUCTIONS FOR 

“JA SERIES” / “JA SERIES PRO” CONTROLLERS

This manual contains safety information that if ignored can 
endanger life or result in serious injury. They are indicated 
by this icon.

Keep the controller protected from sun and water. 
Avoid water splashes.

Read Carefully !

ENGLISH Version

R19-07-19

Summary of Contents for JA PRO Series

Page 1: ...NTROLLERS This manual contains safety information that if ignored can endanger life or result in serious injury They are indicated by this icon Keep the controller protected from sun and water Avoid w...

Page 2: ...authorized personnel only Always discharge the liquid end before servicing the controller Empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals...

Page 3: ...45 plastic box INPUTS Stand by Flow Conductivity probe Temperature probe Ohm H3 PO4 H2 SO4 HNO3 HCl NaOH NaCl TDS PPM CaCO3 and NaCl Some functions are available for PRO model only see page 25 The whe...

Page 4: ...9 10 Standby contact 11 12 Opto coupled output NPN PULSE To use with IS MF PLUS series pumps max 50mA 24VDC 13 GND 14 Blue 15 Black 16 Brown Proximitysensormod SEPR dont removejumperonblocks13and14 17...

Page 5: ...max 50mA 24VDC 13 GND 14 Blue 15 Black 16 Brown Proximitysensormod SEPR dont removejumperonblocks13and14 17 RS485 18 RS485 RS485 Modbus Communication 20 Black Blue 21 Grey or Green 22 Red Inductivity...

Page 6: ...isable the controller it still remains powered These fields are related to STANDBY NO FLOW ALARM condition During critical situations a warning alarm message may appear To in depth explanation complet...

Page 7: ...s status Software Release Local Time Local Date Setpoint Output 1 status Setpoint Output 2 status Pulse Output status YES ENABLED mA Output reading LAN connection status option WIFI connection status...

Page 8: ...b to save the settings The new administrator password ADMIN is now active To set a new passcode choose GENERAL OPTIONS from Main Menu move on PASSWORD click on wheel and enter a four numbers code Move...

Page 9: ...o scroll through all the options available Settings see page 9 Flow see page 20 General Options see page 20 Reset Controller restore all parameters Out mA see page 21 Standby see page 21 Measure see p...

Page 10: ...nge item is replaced by the working scale selector ppm TDS 3000 mS 300 mS 30 mS Note these changes must be set according to probe reading capacity See probe s datasheet to choose correct values Settin...

Page 11: ...t 100uS ON and 90uS OFF The difference between the two Conductivity values is called HYSTERESIS Setpoint output will stay OFF until reading value will reach 100uS Setpoint output will stay ON until re...

Page 12: ...able the Conductivity pump To use this mode move cursor on Working Mode Press the wheel and select it PROPORTIONAL MODE between 10mS 0 and 25mS 100 100 seconds scale This mode operates Conductivity pu...

Page 13: ...a control element such as the position of a control valve a damper or the power supplied to a heating element Before enabling this mode configure the required parameters within SETTINGS PID PARAMETERS...

Page 14: ...s second point calibration value move the cursor to check to confirm the displayed value To stop the calibration procedure select X Temperature calibration A professional thermometer is required to ob...

Page 15: ...sensor on it Wait until reading value is stable and accor ding to buffer solution value move wheel until it is the same on display Cal at field Default value is 7 00pH To end procedure move cursor on...

Page 16: ...nt of the pumps group The choice of method is exclusively at the user s discretion In both cases to establish the set point value a check using the DPD1 or other analysis system is necessary The enclo...

Page 17: ...mperature compensation Alpha can be changed within 0 0 and 5 0 It should be set according to measured chemical properties If a temperature probe is installed choose Enable for Automatic Otherwise sele...

Page 18: ...t be dry and clean and not installed in plant Use chlorine free water or a carbon filter system and dip probe s head into it then ap on First Point zero to confirm it For Second point calibration use...

Page 19: ...become a de facto standard communication protocol and it is now a commonly available means of connecting industrial electronic devices Set the communication speed according to the PLC system availabl...

Page 20: ...ified amount of time 0 to 59 seconds the controller switches back to normal activities GENERAL OPTIONS Several parameters can be set using this menu such as date time language display contrast main me...

Page 21: ...g procedure using a N O contact mode normally open or N C contact mode normally closed when status on blocks changes Rotate wheel to choose between DISABLE REVERSE N O contact or DIRECT N C contact Fu...

Page 22: ...del 999 999 mV Conductivity range for JA RH mod BS0BN00200002 2000 2000 mV Chlorine range for JA CL model based on installed probe TDS PPM CaCo3 and NaCl 0 001 9999 ppm Temperature range for JA TEMP m...

Page 23: ...00 800 000 700 000 600 000 500 000 400 000 300 000 200 000 100 000 0 5 10 15 20 25 30 35 40 45 50 55 60 CONDUCTIVITY mS CONCENTRATION NITRIC ACID HNO3 SULFURIC ACID H2 SO4 HYDROCHLORIC ACID HClO POTAS...

Page 24: ...4 5 6 A A B B C C D D Progettato da Controllato da Data Edizione Foglio Strumento_JA_Datasheet Massimo_F Massimo_F 29 09 2017 00 Toll Gen 0 05 Peso Lordo Peso Netto Materiale 1 1 96 20 96 20 87 10 73...

Page 25: ...MIN ENABLED ON ALARM CONTACT DELAY RECOVERY YES scroll scroll scroll scroll scroll E E E E E scroll scroll scroll scroll scroll E E E Main Menu E Automatic Ohm TDS PPM 0 999 9nS 0 999 9uS 0 999mS scr...

Page 26: ...26 Motorized electrovalve connection Phase Neutral 230VAC 230VAC SetPoint 1 Output Input...

Page 27: ...heck 7 Password 8 Main Menu 9 Settings Measure JA CD 10 Settings Set Point 10 ON OFF mode 11 Proportional mode 12 PID 13 Calibration JA CD model 14 Calibration JA PH model 15 Calibration JA RH model 1...

Page 28: ...ial types and send them according to local recycling disposal requirements We appreciate your efforts in supporting your local Recycle Environmental Program Working together we ll form an active union...

Reviews: