SETUP

The instrument shows:

1a) ->OFF

  1.00Cl

Pressing twice “RIGHT” key the display will show:

1b) ->ON

  0.00Cl

“1a)” and “1b)” are working range values. Every range is set by a value that operates relay output. Using 
default values, the setpoint 1 will drive a pump that will increase the conductivity value. The pump will 
switch off for Cl minor than 1.00 Cl and switch on when reaches 0,00 Cl: this is the ON/OFF working 

mode. The difference between the two Cl (1.00 and 0.00) is called HYSTERESIS.

To avoid damaging the relay do not set Hysteresis under 0.20Cl. 

PROPORTIONAL mode. The output relay is modulated (ON/OFF) depending on Cl values set in 1a) 

and 1b). Ex.: Proportional mode between 1.00Cl (0%) and 0.50Cl (100%). In this way the relay will be 

on for values between 0.00Cl and 0.50Cl, it will be on for values higher than 1.00Cl. For values between 

1.00Cl and 0.50Cl the relay will be on or off depending on calculated percentage.

The calculation is based on a 100 seconds time. If Cl reaches 0.60Cl  the relay will be on for 80 seconds 

and off for 20 seconds. To set the PROPORTIONAL functioning mode substitute percentual value to 

“ON” and “OFF”.

It is not possible to enter mixed values (ex.: 100% for SP1 and ON for SP2). In this case it will not 

possible save them.

In proportional mode the instrument  doesn’t show  % values for setpoints status but only if they  

are on or off.

It is possible modify setpoint status  (on,off, %) and setpoint values  with RIGHT key on the function to 

modify and using  “UP” and “DOWN” key to change the value. Enter the value and save by pressing  

“E” to save. 

Repeat sequence for setpoint 2 (OUT2).

Press “E” key to exit from setpoint setting. The display will show “SetPoint Saved” and it return 

to the previous menu.

Out mA

This function allows to program the current output for driving a metering pump or a chart recorder.It is 
possible to program in 0

÷

20mA mode or 4

÷

20mA mode in the working range between 0 and 9.99ppm. 

The current output and the working range may be modified. Ex.: it is possible to set 4mA at 9.99ppm 

and 20mA at 0ppm. In this case, if you connect a metering pump it will dose at the maximum flow at 

0ppm. The pump will reduce the flow when Cl decrease and it will stop at 9.99ppm. If you connect a 

chart recorder, a graphic curve will be printed. Set minimum current value to 0ppm and maximum current 

value to 9.99ppm. Use “UP” and “DOWN” keys to change these settings. Press “RIGHT” key to scroll 
through the options. Press “E” to confirm.

Summary of Contents for J DIGITAL CL

Page 1: ...ns important safety informations about installation and use of this equipment Ignoring this informations could result in injuries or damages It is strictly forbidden to use this equipment with radioac...

Page 2: ...authorized personnel only Always discharge the liquid end before servicing the instrument Empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals A...

Page 3: ...VIEW ON MENU 6 SETTING SETPOINTS 1 SETPN 6 Out mA 7 CALIBRATION 2 CALIB 8 DELAY 3 Delay 9 RESERVED 4 Reserv 9 PASSWORD SETUP 1 New Pw 10 STANDBY SETUP 2 Stand 10 MANUAL WORKING MODE 3 Serv 11 ERROR ME...

Page 4: ...red data The Enter key confirms the selection Press RIGHT key for about 4 seconds to switch off the instrument the display shows OFF Repeat the sequence to return to the normal operating mode To reset...

Page 5: ...instrument is set to work without a proximity sensor If needed proximity sensor turns the instrument into stand by mode if there isn t flow Instrument s display shows NO FLOW Connect SEPR to blocks 13...

Page 6: ...MAIN menu is configured as follows 1 Setup This menu is divided into 1 Setpn 2 Calib 3 Delay 2 Param This menu is divided into 1 New Pw 2 Stand 3 Serv service mode SETTING SETPOINTS 1 SETPN Out 1 and...

Page 7: ...and OFF It is not possible to enter mixed values ex 100 for SP1 and ON for SP2 In this case it will not possible save them In proportional mode the instrument doesn t show values for setpoints status...

Page 8: ...top probe connector or if not present disconnect probe s green wire signal wire Wait until R value is stable ATTENTION IT IS NOT NECESSARY THAT R VALUE WILL BE THE SAME OF C VALUE Press E to confirm...

Page 9: ...aches the setpoint values Default value is set to 0 Select DELAY from SETUP menu The instrument will show OUT1 10 Sec Press RIGHT key The instrument will show OUT2 5 Sec Delay time can be set from 0 n...

Page 10: ...Press E to confirm The instrument will show the new password for 2 seconds and then will return to the main menu STANDBY SETUP 2 Stand For a good electrodes polarization a delayed startup every time t...

Page 11: ...tact closed and N C contact open OFF is the status of N O contact open and N C contact closed ERROR MESSAGES If the display shows an error message use the following table ERROR 1 Buffer solution error...

Page 12: ...on center for the recycling of electronic and electrical equipment For more information visit the online site When dismantling a pump please separate material types and send them according to local re...

Reviews: