6
VIEW ON MENU
Make connections and plug the instrument. Instrument’s version message on display confirms that the instrument is on. The
display shows also the probe reading status and if setpoints are active. Press “E” for about 4 seconds to enter into setup menu.
Note: During setup the instrument is in “Stand-by” mode. If user doesn’t press any key for one minute the instrument will
return to normal working mode.
To enter into setup menu, a password is needed. If this the first time that user enters into this menu or if the password is not
changed, simply press “E”. Use “UP” and “DOWN” keys to scroll.
“MAIN” menu is configured as follows:
1)Setup
: This menu is divided into
1)Setpn
;
2)Calib
;
3)Delay
.
2)Param
: This menu is divided into
1)New Pw
;
2)Stand
.
3)Serv.
: service mode.
SETTING SETPOINTS: 1)SETPN
“Out 1” and “Out 2” are relay’s driven outputs. The instrument can be programmed to operate in “On/Off” mode or
“Proportional” mode.
“Proportional” mode can be selected in % between 0 and 100.
Select “SETUP” from “MAIN” menu. Select “SETPN” and press “E”. The display shows:
Setpoint
1) Out 1
Pressing “UP” or “DOWN” key the display will show in succession:
Setpoint
1) Out 1
Setpoint
2) Out 2
Setpoint
3) Out mA
Select the setpoint to set (Out 1 or Out 2) using “UP” and “DOWN” keys, press “E” to confirm. “Out mA” allows to set the
current output based on two Conductivity values. In the next example setpoint 1 (Out 1) has been set.
Installation