5.5.4 Test set-up for table top equipment
Test set up
Test sequence
TRANSIENT-2000
1
CNEFT
GRP
RP / HCP
GRP = Ground reference plane ESD
RP = Reference plane EFT
HCP = Horizontal coupling plane ESD
2x470k
CNEFT = Coupling clamp for EFT
EUT1
EUT2
Isolation
Isolation
I. EFT
1. Connect the earth bar of the TRA3000 F-S-D-V-
C with the flat multiwire cable (1) to the
reference ground plate
2. Put 10 cm insulation between EUT and the
reference ground plate
3. Carry out the tests!
II. ESD
1. Remove the flat multiwire cable (1) between the
earth bar of the TRA3000 F-S-D-V-C and the
reference ground plate
2. Put 0,5 mm insulation between EUT and the
reference ground plate
3. Carry out the tests!
III. SURGE, DIPS, VARIATION
1. Reinstall the flat multiwire cable 1
2. Carry out the tests!
CDN-1000
1
GRP
RP / HCP
GRP = Ground reference plane ESD
RP = Reference plane EFT
HCP = Horizontal couplinge plane ESD
TRANSIENT2000
2x470k
EUT
I. EFT
1. As for single phase EUT
2. As for single phase EUT
3. Connect the Impulse out of the TRANSIENT
with EFT coupling on the threephase
Coupling/De-coupling network CDN-2000-06-32
4. Carry out the tests!
II. ESD
1. As for single phase EUT
2. As for single phase EUT
3. Carry out the tests!
III. SURGE,
1. Make connection 1
2. Connect the surged phase for synchronisation
with EUT Power 1
3. Carry out the tests!
IV. DIPS Interruption
1. Loop the phase for dips and interruption
through the TRA3000 F-S-D-V-C (EUT Power
1)
2. Carry out the tests!
TRANSIENT-2000
1
GRP
RP / HCP
GRP = Ground reference palne ESD
RP = Reference plane EFT
HCP = Horizontal coupling plane ESD
HARMONICS-1000
2x470k
EUT
I. TRA3000 F-S-D-V-C Tests:
1. Carry out the tests as explained for single and
three phase EUT
II. HARMONICS-1000-Measurements
1. Harmonics in accordance with IEC 61000-3-2
2. Flicker in accordance with IEC 61000-3-3
3. Immunity Harmonics IEC 61000-4-13
For brochures and further information about
HARMONICS-1000 contact EMC PARTNER AG or
your nearest representative.
Single Phase EUT
Three Phase EUT
Combination: Immunity and Emission
Summary of Contents for EXTTRA3000 E
Page 8: ...TRANSIENT3000 8 162...
Page 102: ...TRANSIENT3000 102 162...
Page 106: ...TRANSIENT3000 106 162...
Page 116: ...TRANSIENT 2000 116 162...
Page 118: ...TRANSIENT 2000 118 162...
Page 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Page 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 150: ...12BAppendix and Corrections 150 162...
Page 156: ...14BIndex 156 162...
Page 158: ...14BIndex 158 162...
Page 160: ...14BIndex 160 162...
Page 162: ...14BIndex 162 162...