12BAppendix and Corrections
144/162
13.1.2 Definition of the ESD Waveform
IEC 61000-4-2 Ed.2
Level
1
2
3
4
Test
voltage
kV+-5%
2
4
6
8
Peak
current
A+-15%
7,5
15
22,5
30
Amplitude
at
30ns
A+-30%
4
8
12
16
Amplitude
at
60 ns
A+-30%
2
4
6
8
It is only possible to check the impulse current by using very expensive measuring equipment. The price of
such an instrument today lies at approx. 50 k$. In addition, persons who carry out such tests must have
some experience with high voltage and high frequency test work, so that they can interpret the measured
values. The calibration and verification of the generators must be carried out by the manufacturer or the
official calibration authorities.
wall cage
2 Ohm Target
50 Ohm
Termination
Attenuator
e.g. 20 dB
Oscillograph
50 Ohm
u2
u1
ESD-Mouse
IEC 61000-4-2 Ed.1
The ESD-current produces on 2 Ohm Shunt a voltage drop u1. The 2 Ohm target is terminated with 50
Ohm to avoid reflection. With the 20 dB attenuator the 60V trop on the 2 Ohm will be reduced to the
allowed input voltage of the oscilloscope. The memory oscilloscope must have a minimum bandwidth of 1
GHz. For all four levels (2,4,6,8 kV) the current wave-form must be within the tolerances as specified in the
IEC standard 61000-4-2.
IEC 61000-4-2 Ed.2
The ESD-current produces on 2 Ohm Shunt a voltage drop u1. The 2 Ohm target is no longer terminated
with 50 Ohm. To avoid reflection the attenuator must be inserted on the target side. With the 20 dB
attenuator the 60V trop on the 2 Ohm will be reduced to the allowed input voltage of the oscilloscope. The
memory oscilloscope must have a minimum bandwidth of 2.5 GHz. For all four levels (2,4,6,8 kV) the
current wave-form must be within the tolerances as specified in the IEC standard 61000-4-2 Ed.2.
Current peak
Summary of Contents for EXTTRA3000 E
Page 8: ...TRANSIENT3000 8 162...
Page 102: ...TRANSIENT3000 102 162...
Page 106: ...TRANSIENT3000 106 162...
Page 116: ...TRANSIENT 2000 116 162...
Page 118: ...TRANSIENT 2000 118 162...
Page 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Page 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 150: ...12BAppendix and Corrections 150 162...
Page 156: ...14BIndex 156 162...
Page 158: ...14BIndex 158 162...
Page 160: ...14BIndex 160 162...
Page 162: ...14BIndex 162 162...