emaux 8552 Operation Manual Download Page 3

P2    Product Over view

Congratulations on your purchase of this pen type ORP 

meter.  Please  read  the  manual  completely  before  using 

this  meter.  Filing  and  keeping  the  manual  for  future 

reference.  Recommended  to  soak  the  electrode  for  at 

least 30 min. before using to clear up the lazy effect or 

to make the ORP electrode wet.

1. Product Overview

1.1 Features

(1) IP65 Waterproof housing.

(2) Big display for easy reading.

(3) Data hold to freeze readings. Pen size, easy to 

fit in pocket. Low batter y indicator.

(4) Pen size, easy to fit in pocket.

(5) Low batter y indicator.

(6) Auto power off.

(7) Power by 4pcs LR44 batteries.

1.2 Material Supplied 

This package contains: 

The meter x 1

LR44 button batter y x 4 

Operation manual

Color box or plain box

2. POWER SUPPLY

The meter is powered by 4 pcs LR44 batteries. 

To check the batter y when:

1. First time use

2. The batter y symbol appears on LCD

3. The meter cannot power on

To install the batteries:

1. Turn off the meter.
2.  Loose  the  batter y  cover  in  counter-  clockwise   

direction. (DON'T discard the black washer!)

3.  Replace  the  old  batteries  with  four  new  button 

cells LR44.

4.  Make  sure  the  batteries  are  in  place  and  the 

polarity is correct.

5.  Put  back  the  batter y  cover  &  washer  and  turn  it 

tightly in clockwise direction

P5    TROUBLESHOOTING

ORP electrode cleaning: 
If sensing element got contaminated, it will result in 
slow  response  and  inaccu rate  reading.  Clean  it  as 
following:

(1) If contamination is a mineral matter, put sensing 

element in HCL solution 0.1N for 10 minutes and 
then rinse it with distilled water.

(2)  If  the  contamination  is  oil  or  grease  coating, 

clean  sensing  element  with  detergent  and  rinse 
with distilled water.

7. TROUBLESHOOTING

Power on but no display

1. Make sure the time of pressing power key is more 

than 1 second.

2.  Check  the  batter y  are  in  place  and  make  good 

contact and correct polarity.

3. Replace with new batteries and tr y again.

4. Move away the batteries for one minute and then 

put back again.

Slow response
Clean  probe  by  immersing  the  electrode  in  tap 
water for 10-15 minutes, then rinse thoroughly with 
distilled  water  or  use  a  general-purpose  electrode 
cleaner.

LCD display E02, E03
Out  of  ORP  measuring  range.  Check  in  normal 
condition first, if remaining showing E02 or E03, the 
electrode is out of order.

8. 

SPECIFICATION

Accuracy      ±
Power Saving
Measurement range
Resolution 
Hold Data
Waterproof
Size

4mV
Auto Power Off
-999mV ~ +999mV
1mV

IP65
165mm(L)x35mm(W)x32mm(T)