EM Test
dito
Control
Short form Introduction
V 2.04
11
/
15
5.1
Connect the
dito
simulator via RS232-Interface
Connection between
dito
and computer is done by RS232 interface via optical link.
The following steps are necessary :
Connect the RS232 optical link to
dito
.
Switch
dito
on
or
go
to
the
start
menu
(double
click on
dito
tri
g
ger button).
In case that
no
communication
is
possible
it
is
recommended
to
go
back
to
the
start
menu
or
switch
on/off
the
dito
.
This
is
not
a
failure
but
a wanted proc
e
dure.
Connect
the
IFA
interface
adapter
to
the
RS232
connector of the computer.
Start
dito
Control (see above).
Set
dito
into remote via software.
The initial status of the
dito
is local. It can be set
into remote by clicking onto the state button in the
status bar and selection of Remote.
It
can
also
be
set
into
remote
by
selecting
the
menu <Wo
rking><State><Remote>.
It is not necessary to set the baud rate, because
dito
Co
n
trol checks the RS232 parameter.
If
for
the
first
time
no
communication
is
possible
the following dialog will appear.
Select
know
the
button
Detect
to
start
the
RS232
Wi
zard.
The
RS232
Wizard
now
automatically
scans
for
the connected
dito
.
At
the
end
of
the
wizard
dito
should
be
detected.
If not, please check all points above.
In case that communication works,
dito
is in Remote and
dito
Control shows the status.
dito
Control can simulate the
dito
simulator. For this mode please select <
Working
><
State
> <
Simulation
>.