EN
10
Technical Specifications of the SonoTT
TM
Clamp-On Transducer
Size (HxWxD), Weight (incl. cable & plug)
25 x 33 x 45 mm, 137 g
(IDs 13479; 13481; 13475; 13478; 13480; 13476)
27 x 38 x 51 mm, 153 g (ID 13477)
Housing and Lid Material
Epoxy Resin, Aluminum, Brass
Cable length
2.9 m
±
5 cm
Connector plug
15-pin high density D-Sub
Safety Class
Type BF applied part
(in combination with the SonoTT
TM
Ultrasonic FlowComputer)
IP-Code
without connector: IP67 (protected against splashing water)
Recommended tube type
flexible, non-reinforced tubing; e.g. silicone, PVC
Acoustic output signal
Frequency 2,25 MHz, p
r
< 1 MPa, I
spta
< 20 mW/cm²,
I
ob
< 100 mW/cm²,
ΔT < 0.3 °C, TI < 1.0, MI < 1.0
Possible adjustment temperature range
4 °C to 41 °C
Expected product life:
10 years
Accuracy
1
in Combination with the SonoTT
TM
Ultrasonic FlowComputer:
Zero flow deviation:
±15 ml/min
Flow Accuracy
2
0 < flow < Qmin
not defined
Qmin to 1000 ml/min
±100 ml/min
> 1000 ml/min
±7 % of the value ±30 ml/min
1
The accuracy stated above can only be guaranteed if the parameters throughout the measurement are the same as
the ones the sensor was adjusted and calibrated for (e.g. medium type, medium temperature, tube size, tube material).
2
Please note:
–
Qmin refers to the minimum flow value for which the accuracy is specified.
–
Qmax refers to the maximum flow value for which the accuracy is specified.
–
The given values describe the positive flow range, but are the same for negative flow values.
Transport and Storage Conditions:
Atmospheric Pressure
70 kPa to 106 kPa
Temperature Range
-20 °C to 55 °C (-4 °F to 131 °F)
Relative Humidity
10 % to 96 % (non-condensing)
Ambient Conditions during Operation:
Atmospheric Pressure
70 kPa to 106 kPa
Temperature Range
10 °C to 40 °C (50 °F to 104 °F)
Relative Humidity
10 % to 96 % (non-condensing)
SonoTT
TM
Clamp-On Transducer
Copyright em-tec GmbH
D162-700 SonoTT
TM
Clamp-On Transducer - Operating Instructions V3.0
Page 23
Summary of Contents for SonoTT
Page 1: ...SonoTTTM Clamp On Transducer Operating Instructions...
Page 3: ...EN Table of Contents English 5 Deutsch 26 DE SonoTTTM Clamp On Transducer...
Page 50: ......
Page 51: ......
Page 52: ......