page 56 of 75
8.5
Recommended Periodic Checks
It is recommended that periodic inspections are carried out at least every 24
months to make sure the device complies with the safety requirements and
delivers the required measuring accuracy, taking into account the valid
international and national standards and regulations. Use of the following
inspection record is recommended by a qualified person who has adequate
training, knowledge, and practical experience to perform these tests.
If the checks reveal faults or limits are exceeded, stop using the device and
contact our Service department (see chapter SUPPORT AND SERVICE)
General
Test organisation,
Name of Tester
Initial test (reference value):
Recurrent test:
Test after repair:
Device
BioProTT™
Type
Reference no.
SN
Accessories
Visual Inspection
Visual check for mechanical damage
that could impair functions
.
Visual inspection
Passed?
Device & accessories configured according this operating instructions and not changed?
Yes
No
Power supply cord
Yes
No
Other connection cables
Yes
No
Housing
Yes
No
Soiling
Yes
No
Is the type plate legible?
Yes
No
Are the safety relevant markings legible?
Yes
No
Are the accessories used compatible?
Yes
No
Checking Functionality according the Operating Instructions
Switch the device on, connect a flow sensor and check the following points:
Function check
Passed?
Does the LED light up (BioProTT FlowTrack and BioProTT FlowTrack DINrail) or the
display light up when the device is switched on?
Is the brightness increasable and reducible? (see chapter
Yes
No
Perform a function check in accordance with the Operating Instructions
(see chapter
PERFORMING MEASUREMENTS ON TUBING SYSTEMS
USING
BIOPROTTTM CLAMP-ON TRANSDUCERS
Yes
No
Clamp-On Transducer: Check coupling value with tube filled:
The acoustic coupling value must be sufficient (see chapter
).
Yes
No