elysator PUROTAP 1000 Installation Function Operation Service Download Page 3

Cleaning system with auxiliary pump

Use of the LF-M Meter

Directly filling via the ion exchanger is not suitable for surface heating systems that can only be de-aerated 
through purging. The delivery rate is insufficient enough for purging air from a horizontal line. In these 
cases, we recommend filling up with untreated water and subsequently demineralising using a cleaning 
system.
This procedure is also suitable for subsequently demineralising systems with too much salt content accor-
ding to the boiler manufacturer's specifications or guidelines. 
In doing so, the ion exchanger is integrated into the heating system circulation with the help of a separate 
pump (e.g. jet pump, impeller pump, centrifugal pump) and two reinforced hoses. The type of adapter 
used here is of less importance; instead, it is more important that the circulation pumps are operating and 
all valves are open enabling the system water to be mixed thoroughly. 

The temperature of the cartridge may be a maximum of 60 °C for a short period.‎ It must be connec-
ted to a return line with the lowest possible temperature. Only hoses that are sufficiently pressure 
and temperature resistant should be used (reinforced hoses). The work may only take place with 
monitoring.

The pump must be connected behind the cartridge, no pump pressure on the cartridge!

General safety instructions

The resin from the cartridge must not enter the heating system. Before each use, the retention sieves 
on both sides of the cartridge must be checked. The work with the cartridge may only take place with 
monitoring. 
It is recommended to do a test run of the system and purge the oxygen and carbon dioxide gases that 
are normally present in tap water.

Run the cartridge until it is exhausted. 
Determine the number or size of the 
required cartridges based on the sys-
tem content and the capacity diagram 
on page 2.

The adhesive label stick visibly to the 
boiler. 
Affix the correct fill-up.

The use of measuring devices makes 
reliable system cleaning easier.

The LF-M meter enables flow rate and 
water quality monitoring. The cartrid-
ge  can  be  completely  used  up  and 
then  replaced  at  the  appropriate 
time.

The cartridge can easily be used seve-
ral  times  with  the  LF-M  meter  for 
smaller  amounts  of  water  until  it  is 
exhausted.

Auxiliary pump

M a i n  d i s t r i b u t i o n

3  EN

LF-M Meter

Reviews: