Maintenance
47
Fig. 23: Replacement of moisture trap
1.
Turn both screws (
1
) counter clockwise, until the ring (
3
) touches the part (
2
).
2.
Lift the glass tube (
5
) upwards, tilt it to the side and pull downwards.
3.
Replace the magnesium perchlorate (Anhydrone). See chapter
reagent tube replacing
4.
Install the glass tube (
5
) in reverse order after greasing the lower O-ring.
5.
Turn both screws (
1
) clockwise until the O-ring (
4
) is properly pressed on the
inner surface of the glass tube (
5
).
Remark
The O-ring imprint on the whole circle length of the glass tube should be of about 2 mm width.
6.3 Reagent tubes filling
6.3.1 Chemicals
CAUTION
C13.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
Parts of the analyzer can be very hot.
•
Use heat protecting gloves.
WARNING
W9.0017
Danger of toxication and personal injuries
Some chemicals may cause a fatal toxication or dangerous skin corrosion.
•
Refer to the material safety data sheet of the used substances.
•
Never eat or drink close to the chemical substances.
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Page 5: ...5 ...
Page 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Page 80: ...Function description 80 Fig 51 Gas flow system ...
Page 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Page 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...