6 Operation
This chapter describes how the Ultrasonic Flowmeter Series 6 can be operated through the
interactive touch screen display on the meter. For more detail information, see the Q.Sonic Series 6
operation and maintenance manual (document code: 22.100.200.002.02/2 last valid revision).
6.1
LED at display
2 LED’s provide an overall status indication of the meter.
•
Power LED (LED 1 at Figure 16):
LED status
Description
Off
Power off
Green
Power on
Table 1: Power LED
•
Status LED (LED 2 at Figure 16)
LED status
Description
Off
Power off or
The device is running warning and error free.
Yellow, flashing
A warning is pending that does not affect the metrology
relevant functionalities. The user has to accept such
warnings at the operation panel before the color of the
status LED turns back to off*
Yellow, permanently
illuminated
At least one warning that does not affect the metrology
relevant functionalities was pending but has already ended.
The user has to accept such warnings at the operation
panel before the color of the status LED turns back to off*
Red, flashing
An error is pending that affects the metrology relevant
functionalities. The user has to accept such warnings at the
operation panel before the color of the status LED turns
back to off*
Red, permanently
illuminated
At least one error that affects the metrology relevant
functionalities was pending but has already ended. The
user has to accept such warnings at the operation panel
before the color of the status LED turns back to off*
* See chapter 6.2.3
Table 2: Status LED