Elster Q.Sonic PLUS Quick Start Manual Download Page 1

 

Elster NV/SA 
Rijkmakerlaan 9 
2910 Essen, 
Belgium 
T + 32 3 670 0700 

www.elster-instromet.com

 

DOCUMENT TITLE 

ULTRASONIC FLOWMETER SERIES 6, MODELS Q.SONIC

PLUS

 

 

SUBTITLE 

QUICK START MANUAL 

DOCUMENT NUMBER 

22.100.200.000.02/2 

REVISION

 

REVISION DATE 

2013-02-22 

 

This reproduction is sole property of Elster NV/SA and is subject to the conditions that it or any information contained therein will not be used in 
any way detrimental to our interests and that all copies will be returned immediately on demand 

PAGE: 

1 OF 29 

 

 

22.100.200.000.02_2_B ULTRASONIC FLOWMETER SERIES 6.DOC 

 

Ultrasonic Flowmeter Series 6

Q.Sonic

plus

  

 

 

Quick start manual

 

 
 
 
 
 
 
 
 

Summary of Contents for Q.Sonic PLUS

Page 1: ...2 2 REVISION B REVISION DATE 2013 02 22 This reproduction is sole property of Elster NV SA and is subject to the conditions that it or any information contained therein will not be used in any way detrimental to our interests and that all copies will be returned immediately on demand PAGE 1 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC Ultrasonic Flowmeter Series 6 Q Sonicplus Quic...

Page 2: ...es 6 gas flow meter 6 2 1 General 6 2 2 Applicable standards 6 2 3 Configuration 6 2 4 Calibration 7 3 Theory of operation 8 3 1 Flow velocity measurement 8 3 2 Correction after calibration 9 3 3 Volume flow at line conditions 9 4 System description 10 4 1 Flow cell 10 4 2 Signal processing unit 10 4 3 Transducers 11 4 4 Flow cell optional pressure sensor 11 4 5 Flow cell optional temperature sens...

Page 3: ...change 25 7 2 4 SPU exchange 25 8 Storage and shipping 27 9 MID requirements 28 9 1 General 28 9 2 EC declaration of conformity 28 9 3 Sealing 28 9 4 Calibration 28 9 5 Installation requirements 28 Appendix I References 29 Figures Figure 1 Path types 6 Figure 2 path layout Q Sonic plus 7 Figure 3 Ultrasonic measuring line 8 Figure 4 Example of an Elster Instromet Ultrasonic gas flowmeter 10 Figure...

Page 4: ...mage before start up during operation and to ensure trouble free operation in the safest possible way Before using the product read this manual carefully familiarise yourself with the operation of the product and strictly follow the instructions If you have any questions or need further details of specific matters concerning this product please do not hesitate to contact one of our staff members e...

Page 5: ...ning A warning indicates hazards or unsafe practices that could result in severe personal injury or death Caution A caution indicates hazards or unsafe practices that could result in minor personal injury or product or property damage A caution is also used to indicate operations or practices that may cause the product to operate in an undefined or unexpected way or may produce non specification r...

Page 6: ...dy transfer measuring applications that demand a high degree of accuracy and reliability 2 2 Applicable standards The Ultrasonic Flowmeter Series 6 is manufactured to be in accordance with European Directives ATEX PED EMC and optionally MID If the meter is ordered for use at a location where European Directives are NOT mandatory the meter can alternatively be manufactured with IECEx or FM approval...

Page 7: ... returned immediately on demand PAGE 7 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC Figure 2 path layout Q Sonicplus 2 4 Calibration When using the Ultrasonic Flowmeter Series 6 model Q Sonic plus in custody transfer applications most countries demand by law a calibration from a certified calibration institute supervised by an inspector of weights and measures Examples of faciliti...

Page 8: ...eries 6 employs reflection paths where the acoustic pulses reflect one or more times off the pipe wall v B A L D ϕ Figure 3 Ultrasonic measuring line 3 1 Flow velocity measurement The acoustic pulses are crossing the pipe like a ferryman crossing a river Without flow they propagate with the same speed in both directions If the gas in the pipe has a flow velocity different from zero pulses travelli...

Page 9: ...FLOWMETER SERIES 6 DOC 3 2 Correction after calibration After flow calibration the meter can be adjusted either through an adjust factor or through linearization How the meter is adjusted can be visualized at the display see chapter 6 1 3 3 Volume flow at line conditions The volume flow at line conditions QLine is the adjusted profile corrected gas velocity Vline multiplied by the internal cross s...

Page 10: ...sors are mounted on the flow cell see Figure 4 Figure 4 Example of an Elster Instromet Ultrasonic gas flowmeter 4 2 Signal processing unit The SPU is mounted in an explosion proof housing The box consists of two separate compartments a main and a rear compartment see Figure 5 The main compartment can be opened from the side of the SPU and contains the main circuit boards The main compartment also ...

Page 11: ...ment are generated and received by ultrasonic transducers Piezoelectric transducers employ crystals or ceramics that are set into vibration when an alternating voltage is applied to the piezoelectric element The vibrating element generates sound waves in the gas Since the piezoelectric effect is reversible the element will become electrically polarised and produce voltages related to the mechanica...

Page 12: ...for The Reynolds flow profile correction Compensation of the flow cell expansion due to flow cell temperature Caution The temperature sensor is not used for volume conversion 4 6 Labels and nameplates Nameplates and labels are used to identify the product and to provide details on the specific product Together with the product manual it specifies how the product is certified and designed The main ...

Page 13: ...nce DEKRA 11ATEX0170 X Warning Read instruction manual before operating device Figure 10 example ATEX label 4 6 2 IECEx certified The explosion proof housing has following IECEx certification information Classification Ex d ia ia IIB H2 T6 Gb IP66 40 C Tamb 60 C IECEx certificate reference IECE DEK11 0062 X Warning Read instruction manual before operating device Figure 11 example IECEx label 4 6 3...

Page 14: ... 22 100 200 000 02 2 REVISION B REVISION DATE 2013 02 22 This reproduction is sole property of Elster NV SA and is subject to the conditions that it or any information contained therein will not be used in any way detrimental to our interests and that all copies will be returned immediately on demand PAGE 14 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC Figure 12 example FM label ...

Page 15: ... important sealing locations This chapter describes the sealings as required by MID certificate T10335 see also chapter 9 Even if MID is not required it is advised to seal the UFM 4 7 1 Main plate Figure 13 shows how the main plate is sealed to the flow cell Figure 13 sealing main plate 4 7 2 SPU The SPU in the main compartment of the flameproof certified box is sealed on 2 locations By means of t...

Page 16: ...herein will not be used in any way detrimental to our interests and that all copies will be returned immediately on demand PAGE 16 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC If both pins on the switch are up the meter is sealed and the parameter settings of the meter are locked from editing If both pins are down the meter is unsealed and parameter settings can be altered when us...

Page 17: ...ra documentation is required and delivered If documentation is missing contact Elster NV SA or your local agent immediately 5 2 Installation requirements flow cell 5 2 1 Installing meter in pipeline The Ultrasonic Flowmeter Series 6 flow cell transducers and SPU is shipped in a suitable box e g wooden box cardboard box Carefully disassemble the box Remove the transport straps from the flow cell an...

Page 18: ...2 100 200 001 02 2 Please refer to these documents for correct wiring of the Ultrasonic flowmeter series 6 5 4 SPU configuration When the meter is installed and correctly wired it s advisable to read out the parameter settings of the meter If the meter had been calibrated before the parameter set up should be compared to the one of the calibration The parameter set up can be read with the software...

Page 19: ...AGE 19 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC 5 6 Hot commissioning The hot commissioning is in most cases the last test of the UFM and can be witnessed by a representative of the client and if required by an inspector of weights and measures for the official sealing Under this condition there is process gas in the pipe and a flow test is being performed The performance AGC ...

Page 20: ... at Figure 16 LED status Description Off Power off Green Power on Table 1 Power LED Status LED LED 2 at Figure 16 LED status Description Off Power off or The device is running warning and error free Yellow flashing A warning is pending that does not affect the metrology relevant functionalities The user has to accept such warnings at the operation panel before the color of the status LED turns bac...

Page 21: ...interactive touch screen display showing the most important measurements and diagnostics line flow gas velocity speed of sound totalizers It contains a touch screen with 7 touch area s see Figure 17 It s also possible to mimic the interactive touch screen display on your PC To do this you need to connect the ethernet cable at the field terminal board see chapter 5 2 2 Go to the internet browser of...

Page 22: ...conditions that it or any information contained therein will not be used in any way detrimental to our interests and that all copies will be returned immediately on demand PAGE 22 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC Verifing the components with their checksums Verifing the software status of the parameter build up Figure 18 checking software versions and checksum through ...

Page 23: ...C FLOWMETER SERIES 6 DOC 6 2 3 Checking errors and warnings The status LED on the display shows when a warning or error has taken place see chapter 6 1 Through the front panel it is possible to get more information on the error warning see Figure 20 It is also possible to accept the error warning Hereby it will be written in the Logbook of the meter and will no longer influence the status LED on t...

Page 24: ... detailed information about maintenance please refer to operation and maintenance manual doc code 22 100 200 002 02 2 last valid revision 7 2 Exchanging components Different parts of the Ultrasonic Flowmeter Series 6 metering system like transducers electronic boards etc can be exchanged easily The digital programmed pulse shape and pulse identification of the meter is always identical Therefore t...

Page 25: ...ties and accuracy thus the calibration of the Ultrasonic Gas Flow Meter However as the transducers are paired up during production Elster NV SA always recommends changing both transducers of an acoustic path if possible Warning Obey the rules and regulations that apply to hazardous area operations and those with respect to custody transfer regulations sealing Exchanging a transducer can take place...

Page 26: ...erests and that all copies will be returned immediately on demand PAGE 26 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC This will not affect the measuring characteristics and the accuracy and as a consequence the calibration of the Ultrasonic Flowmeter Series 6 However if the board is sealed after calibration please contact Elster NV SA or your local representative before proceedin...

Page 27: ... demand PAGE 27 OF 29 22 100 200 000 02_2_B ULTRASONIC FLOWMETER SERIES 6 DOC 8 Storage and shipping As the Ultrasonic Flowmeter Series 6 is a delicate instrument care should be taken to carefully handle and store the flowmeter in a proper way Improper handling shipping or storing may void its warranty The Ultrasonic Flowmeter Series 6 should be stored in indoor conditions with a low humidity 5 95...

Page 28: ...PED Equipment and Protective systems intended for use in potential Explosive Atmospheres ATEX Directive ElectroMagnetic Compatibility EMC Directive Measuring Instrument Directive MID In compliance with the applicable directives the meters are supplied with the CE mark and the EC Declaration of Conformity This declaration is part of your flow meter documentation since it also contains important det...

Page 29: ...below can be obtained at Elster NV SA 1 MID certificate T10335 last valid version Doc No T10335_certificate 2 Ultrasonic Flowmeter Series 6 operation and maintenance manual Doc No 22 100 200 002 02 2 last valid revision 3 Ultrasonic Flowmeter Series 6 wiring instructions Doc No 03 302 101 003 07 2 last valid revision 4 Transducer exchange at atmospheric conditions Doc No 03 200 001 001 02 2 last v...