EK280
Application Manual
12
Method (without opening the calibration lock):
Open the administrator lock
Move the cursor to
"Serv."
>
"Volume" > "Actual volume" > "Synchronization Vm > "VmA"
(the value
to be taken over, is first interim saved by the "resettable interim counter" VmP and then taken over
as "Vm" when the counter reading is reached).
Press the ENTER button
One digit of the value flashes.
Move the cursor with the arrow buttons
or
to the digits and change these with the arrow buttons
or
.
Press the ENTER button to confirm the value setting. Input can be canceled using the ESC key.
To accept the set value move the cursor to the value
"Store"
Press the ENTER
button
"0"
flashes.
Change the value to
"1"
by pressing the
or
arrow keys.
Press the ENTER button to confirm the value setting. Input can be canceled using the ESC key.
For
"Vm"
is taken
"VmP"
.
i
If the administrator lock is open and the transfer of "VmP" for "Vm" is rejected with the
message
--
13
--
, this process has previously been carried out.
Further changes are possible when either the calibration lock or the administration lock is open,
after deleting the change information regarding setting "Vm" at the same value as the gas meter
(see chapter 2.4.4.1).
The following values are shown on the same menu:
Display
Description
TimeX
Time when counter synchronization was carried out
Vm_o
"Old" operating volume counter value Vm
Vm_n
"New" operating volume counter value Vm
2.4.4.1 Delete change information for synchronization Vm with the gas meter
Change information for adapting "Vb" to the gas counter must be deleted while the calibration lock is open in
order to re-set the volume under metering period conditions when the administrator lock is open:
Open the calibration lock
Move the cursor to
"Serv."
>
"Volume" > "Actual volume" > "Synchronization Vm > "Clear".
Press the ENTER button
"0"
flashes.
Change the value to
"1"
by pressing the
or
arrow keys.
Press the ENTER button to confirm the value setting. Input can be canceled using the ESC key.
"TimeX"
(time stamp of synchronization),
"Vm_o"
(Vm old value),
"Vm_n"
(Vm new value) will be set
to default values.
2.4.5
Parameterize the volume at base conditions
i
A one-off setting of the volume counter is possible from software version 2.20 while the
administrator lock is open.
The quantity converter EK280 offers the possibility of adjusting the standard volume counter once during
commissioning of the unit while the administrator lock is open. Setting the volume is always possible while
the calibration lock is open and can also be carried out directly via Standard Volume (tab
"Serv."
to
"Volume"
> "Standard Volume" > "Vb"
.