![Elster Instromet EnCal 3000 Quad Hardware Manual Download Page 19](http://html.mh-extra.com/html/elster-instromet/encal-3000-quad/encal-3000-quad_hardware-manual_2396892019.webp)
Elster GmbH
___________________________________________________________________________________________________________
EnCal 3000 Quad– Hardware Manual
22/06/2017
19/52
2.5
Processor Board
The main components of the board are:
Arm9 Processor
A Flash-Memory 32 MB for local data storage
The Ethernet-Port
Data communication ports (COM1-4) for serial Modbus, Analogue and Digital I/O
USB Port
I/O Controller for Communication to the analytical channels
Pressure- and Temperature Sensor
fig. 2-5 Processor Board connections
Save parameters in the measuring unit
Setting the "Configuration Jumpers" can protect the parameters on the measuring unit computer witch was
loaded by using the software RGC 3000. This configuration Jumper is on the mainboard (see fig. 2-6). If the
jumper is set, the parameters inside the measuring unit cannot be override.
Changes in the parameters can only be transferred to the measuring unit, if the jumper is not set. The setting
of the jumper can be checked on the display of the “instrument status” page. See Section 3.2 of the software
manual. How to set the jumper is shown in Figure 2.6.
Arm9
Processor
USB
DHCP
Switch
Configuration
Jumper
Digital I/O
COM2
COM1
Ethernet
Port
Analogue
I/O
SD-Card
Serial Ports
COM3 + 4
I/O Controller