ELPRO ECOLOG-PRO Series Operation Manual Download Page 7

Document: EP6001Eb

SYSTEM DESCRIPTION

7 - EN

1.1

Configuration and Evaluation

Configuration, calibration, and monitoring are carried out with the elproMONITOR software.

SM3031D, elproMONITOR operating instructions

1.2

Notes

Power supply unit

Follow the safety and application instructions of the power supply unit. Only operate the 
device using the delivered power supply unit. Use of an incorrect power supply unit may 
cause damage to the device. Always replace defective or damaged power supply units.

Cleaning

To clean the modules, use a slightly moist cloth. Never use thinner, gasoline, alcohol, or 
any other cleaning agents since they may damage the housing.

Environmental 
Conditions

• Make sure the modules are at room temperature before initial startup.

• No guarantee can be given if functions exceed the specified threshold range. Ambient 

temperature 0°C to 40°C, 10% RH to 90% RH, non-condensing

• Protection class IP20

Summary of Contents for ECOLOG-PRO Series

Page 1: ...ECOLOG PRO Modules Operation Manual ...

Page 2: ...s free from material and processing faults under normal conditions All main items of the software product comply with the operating instructions and the information in the help file The guarantee does not apply to consumables disposable batteries or any other product that ELPRO BUCHS AG deems to be misused modified replaced by an incorrect product damaged as the result of an accident or operationa...

Page 3: ...cuit Diagram 11 3 3 Sensor Status 11 4 ECOLOG PRO 4MA 4 20 mA Measurement Module 12 4 1 Circuit Diagram 13 4 2 Sensor Status 13 5 Module with Contact Inputs ECOLOG PRO 4DI 14 5 1 Contact Statuses 15 5 2 Circuit Diagram 15 6 Module with Contact Outputs ECOLOG PRO 4DO 16 6 1 Contact Statuses 17 6 2 Alarm Indicator Power Supply and Ground 18 6 3 Local Alarm 18 6 4 Alarm Forwarding 18 7 1250 mA Power ...

Page 4: ... elproMONITOR SV6031D In the interests of our customers we reserve the right to make any changes result ing from technical advances For this reason diagrams descriptions and the scope of delivery are subject to change without notice This manual is valid as from Version 2017 CHANGES OR MODIFICATIONS MADE TO THIS EQUIPMENT MAY RENDER NULL AND VOID THE ETSI FCC AUTHORIZATION TO OPERATE THIS EQUIPMENT...

Page 5: ...anufacturer guarantees that this product complies with the relevant guidelines EN 61000 6 2 2001 and EN 61000 6 4 2001 FCC ID Z45 E11645398 IC 9954A E11645398 This product must be disposed of in accordance with WEEE Waste Electrical and Electronic Equipment 2002 96 EC ...

Page 6: ...e LBR ECOLOG PRO 4MA is the measurement module for analog signals The module has 4 inputs for standard 4 20 mA signals The monitored transmitters can be powered by the module or by an external power supply For operation it is connected to a LAN bridge LBR ECOLOG PRO 4DI is the module with 4 contact inputs digital inputs for the ECOLOG PRO system The monitored contacts are powered by the module For...

Page 7: ... power supply unit Use of an incorrect power supply unit may cause damage to the device Always replace defective or damaged power supply units Cleaning To clean the modules use a slightly moist cloth Never use thinner gasoline alcohol or any other cleaning agents since they may damage the housing Environmental Conditions Make sure the modules are at room temperature before initial startup No guara...

Page 8: ...atus green link ok red link inactive Power supply 24 VDC 10 to 20 By external power supply unit Maximum connection cable length 30 m Functional ground is optional Maximum number of measure ment modules The maximum number of measurement modules is defined by the specifica tions of the power supply unit or the limits of the ECOLOG PRO LBR Maximum 2500 mA system bus 32 measurement modules Power suppl...

Page 9: ...G PRO LBR COMMUNICATION MODULE 9 EN 2 1 Power Supply Unit Circuit Diagram Power supply unit Chapter 7 1250 mA Power Supply Unit Accessories Functional ground Connect the functional ground when shielded sensor cables are used ...

Page 10: ...Sensor error and cable fault are not considered Measurement resolution 0 1 C Cable fault up to 30 m 4x0 25 mm2 without shield 30 m to 50 m 4x0 25 mm2 with shield 50 m to 80 m 4x0 34 mm2 with shield 0 06 C 0 1 C 0 2 C Logging interval 1 minute to 60 minutes Each channel is capable of logging at a separate interval Date and time Internal system clock The clock is synchronized permanently to elpro MO...

Page 11: ...is stored in the internal non volatile module memory Content of package Measurement module Connecting hook see module rear Information leaflet Validation certificate The plugs for sensor connection are not included in the scope of supply Cable A Cable B Cable C Part No 800508 M12 plug to Pt100 sensor Terminal Cable A Cable B Cable C 1 red white white 2 red brown brown 3 white black green 4 white b...

Page 12: ...4 mA Transmitter power supply 24 VDC 20 Pmax 3 W Load 50 to 100 ohms Measurement accuracy of module 0 04 mA up to 30 m 4x0 25 mm2 without shield Measurement resolution 0 01 mA Connection cable length max 200m Logging interval 1 minute to 60 minutes Each channel is capable of logging at a separate interval Date and time Internal system clock The clock is synchronized permanently to elpro MONITOR sy...

Page 13: ...h external power supply 4 2 Sensor Status Calibration data Data is stored in the internal non volatile module memory Content of package Measurement module Connecting hook see module rear Information leaflet Validation certificate The plugs for sensor connection are not included in the scope of supply Terminal 1 Ground 2 3 Signal 4 5 Power supply Status dark No channel configured green All configur...

Page 14: ...ent for all 4 sensors is 120 mA Statuses Open Closed Connection cable length max 200m Logging interval 1 minute to 60 minutes Each channel is capable of logging at a separate interval Pulse length 200 ms Minimum time interval to detect status change Date and time Internal system clock The clock is synchronized permanently to elpro MONITOR system time Module power supply Via internal system bus Dat...

Page 15: ...in each measurement sampling period A Number of edge changes B Total duty cycle as percentage of measurement sampling period C Present contact status Example 5 2 Circuit Diagram CH1 Example Switch or 2 wire sensor powered by the module CH3 Example 3 wire sensor powered by the module Door monitoring Time interval of open doors Number of door openings per measurement sampling period Terminal 1 Groun...

Page 16: ...DCV see 6 2 Alarm Indicator Power Supply and Ground External power supply unit for alarm indicators 12 to 24 VDC Max switching current per contact 100 mA Connection cable length max 200 m Threshold values and time response The contacts are parameterized threshold values and alarm delay times by the elproMONITOR Date and time for alarm delay times Internal system clock The clock is synchronized per...

Page 17: ...THE POWER SUPPLY UNIT MUST BE CONNECTED TO ENABLE ALARM SIGNALS Pin 1 GND Com contact Pin 2 NO Normally Open contact Pin 3 GND Com contact Pin 4 NO Normally Open contact Pin 5 Contact for external power supply Internal connected to all 4 module channels The NC and NO contacts are open in these operating states No alarm signals are generated No alarm is present in this operating state An alarm is p...

Page 18: ...when the contact closes 6 4 Alarm Forwarding Example Contact CH1 monitored by central plant system An alarm is triggered when the contact opens This circuit layout allows wire break monitor ing Alarm indicator grounds are electrically isolated from the system bus ground Terminal 1 Ground 2 3 Ground 4 Closes in event of an alarm NO 5 Alarm indicator power supply Only one power supply unit is possib...

Page 19: ...e design plug in power supply units are delivered with replaceable country specific plugs for AC input Power supply unit see Accessories for Part No Input voltage 100 240 VAC 10 50 to 60 Hz Country versions Europe U K U S A Output voltage 24 VDC 5 Output current max 1250 mA Dimensions 90x55x34 Connection cable length 1830 mm Content of package Power supply unit with connection cable and plug for c...

Page 20: ...y unit and the network To carry out the configuration the ECOLOG PRO LBR must be in the same logical network as the PC which is used for the configuration Start Module Configu rator as administrator Step 1 In the start window of the Module Configurator left click on the field LBR and open the win dow to select the search functions Search the ECOLOG PRO LBR using one of the 3 methods below Start wi...

Page 21: ... IP address 3 Hostname Define the ECOLOG PRO LBR for configuration by entering the host name Standard Host name serial number Step 3 After a successful search the input fields are completed with the current settings of the selected module and enabled for processing Enter or adapt and assign the network parameters in this window ...

Page 22: ... U S plug 802100 CTR_ECOLOG PRO Channel Plug to connect a sensor 802101 CTR_ECOLOG PRO LBR Plug to connect a power supply unit 800508 CTR_M12F_IDC M12 F cable coupling to connect sensor 800513 CTR_M12M_IDC M12 M cable coupling to connect sensor Author Date Version Description ANGUB March 8 2017EP6001D First release ANGUB 16 June 2017 EP6001Da Designation of power supply unit changed ANGUB 24 Aug 2...

Page 23: ...EN 23 Document EP6001Eb ...

Page 24: ...Operation Manual ECOLOG PRO EP6001Eb ELPRO BUCHS AG Langäulistrasse 45 9470 Buchs SCHWEIZ E Mail swiss elpro com ...