background image

USER MANUAL

Centrifuge-Mixer CM-50MP

24

Phone:

 

(805) 375-0060

 3483 Old Conejo Rd ., STE 206

Newbury Park, CA 91320

info@elminorthamerica .com

elminorthamerica.com

North America

Phone:

 

(371) 6755-8743

21-133 Aizkraukles Str ., Riga LV-1006 

Fax: (+371) 6755 1934

E-mail: info@elmi-tech .com

www.elmi-tech.com

Europe

© ELMI North America, Inc . 2021

Summary of Contents for CM-50MP

Page 1: ...USER MANUAL elminorthamerica com Centrifuge Mixer CM 50MP...

Page 2: ...of the Fugamix 3 10 Operating fugamix in Miniprep mode Understanding the interface 11 Operating fugamix in Miniprep 1 mode 12 Operating fugamix in Miniprep 2 mode 14 Operating fugamix in Custom mode...

Page 3: ...xing level and rotor speed simultaneous RPM and RCF The centrifuge lid locks during rotation and unlocks at the end of each run Adjustable sound alarm brightness and sleep mode are also standard Imbal...

Page 4: ...tors with visible signs of corrosion wear or mechanical damage Do not fill in the containers after they are insert ed in the rotor Do not leave the operating unit unattended Do not operate the unit in...

Page 5: ...5 USER MANUAL Centrifuge Mixer CM 50MP Casing Control panel Lid Network switch Power plug...

Page 6: ...ign allows the user to spin and vigorously shake any substance including heavy beads for as long as the experiment requires Assembly Put the rotor on the shaft first and then tighten the nut with a pr...

Page 7: ...ixer CM 50MP Lid opening Mode selection Start stop and custom program launch Custom Miniprep 1 and Miniprep 2 mode view and configuration Up and Down menu scroll and parameter change Mixer start stop...

Page 8: ...r volume and more When timer is set to Short run you can push and spin the centrifuge Operation will stop when but ton released When timer is set to Short run you can push and mix Operation will stop...

Page 9: ...lor means that the lid is opened Settings are made using the up down keyboard Pa rameters can only be changed while the centrifuge is not running By reducing the timer to the minimum you will reach a...

Page 10: ...ously No need for manual pipetting and vortexing 2 Mixing of lysis and neutralizing buffers is done automatically and simultaneously 3 Plasmid DNA extraction is reproducible rapid and efficient 4 High...

Page 11: ...can disable the spin down part and do only the Mixing Operating fugamix in Miniprep mode Understanding the interface Examples of different steps Miniprep plasmid DNA extraction mode is a se quence of...

Page 12: ...x homogeneously and neutralize Followed by 5 min spin down Pro teins and DNA is now separated Step 1 Load test tubes containing bacteria in to the rotor Press Start Stop and wait 1 min until the cells...

Page 13: ...re then 95 of plasmid is in CCC format Plasmid DNA extracted can be used for se quencing cloning transformation and molecu lar biology studies Ratio OD 260 280 1 85 1 9 Step 8 Move the spin columns to...

Page 14: ...Apply 700 l of washing buffer on the spin column Press Start Stop and spin down for 15 sec Step 1 Add 20 l RNaseA 10mg ml 100 l of resus pension buffer and press Start Stop Two solutions will mix hom...

Page 15: ...plasmid is in CCC format Plasmid DNA extracted can be used for se quencing cloning transformation and molecu lar biology studies Ratio OD 260 280 1 85 1 9 Step 6 Move the spin columns to the elution...

Page 16: ...he spin down part and do only the Mixing Operating fugamix in Custom mode Understanding the interface Way to program individual steps Custom mode is a step by step program with 9 inde pendently progra...

Page 17: ...nd RPM by using up down keyboard When all the steps are edited press SET and return to main screen of the Custom mode You will see your edited steps as a grey cells with white letters n order to start...

Page 18: ...op open the lid and wait to proceed to the next step F2 Steps that are already completed are marked black color If needed you can move selector from one step to another to change the sequence using th...

Page 19: ...0 9 1332 10 1400 11 1540 12 1620 By reducing the timer to the minimum you will reach a Short run mode This mode allows using MIX STOP button to manage the duration of vortexing Vortexer will run while...

Page 20: ...overheated Give a little time and centrifuge will cool down automatically Motor is overheated Give a little time and centrifuge will cool down automatically Rotor detection fails Centrifuge does not s...

Page 21: ...e will enter to a spe cial service menu TO Press and hold down both buttons and While holding down the but tons click the power switch to on position Compare the modifiable parameter figures with corr...

Page 22: ...ixed and transportation runs accordingly to regulations applied to this way of transport Equipment should be stored in original manufacturers packaging in dry room with humidity not more then 80 and t...

Page 23: ...control person______________________________ name signature Date of manufacture _______________________________ PLACE FOR STAMP Certificate of sale Organization______________________ Address 1 line __...

Page 24: ...83 Old Conejo Rd STE 206 Newbury Park CA 91320 info elminorthamerica com elminorthamerica com North America Phone 371 6755 8743 21 133 Aizkraukles Str Riga LV 1006 Fax 371 6755 1934 E mail info elmi t...