Elmi CM-50 User Manual Download Page 5

Preparation to work

•     

Unpack the centrifuge.

•     

Inspect the appearance of the device, the adapter and the power cord for damage.

•     

Open the lid of the centrifuge with the emergency lid opening lever and remove the 

packaging material (see emergency opening of the centrifuge lid pg.7).

•     

In the cold season keep centrifuge at room temperature for at least 2 hours.

•     

Assembly the rotor it must rotate freely without jamming.

•     

Connect the centrifuge to the adapter and the adapter to the socket.

•     

The light indicators will light up.

•     

In the absence of faults or damage the centrifuge can be considered ready to work.

ATTENTION! In the presence of faults and damage, do not use the centrifuge without 

consulting a specialist.

ATTENTION! Network socket must correspond to the adapter plug. In case of non 

compliance with this condition the rules of electrical safety are violated.

4

 

Control description

On the front side of the centrifuge the control panel is situated with indicators and buttons that 
have the following functions:

Control description

- Increasing parameter / lid opening button.

- Decreasing parameter  / starting and stopping button.

- Setting timer.

- Setting braking level.

- Setting speed.

ATTENTION! To change the PRM  to G (RCF), or vice versa,
Press and hold             button for 5 seconds.

Summary of Contents for CM-50

Page 1: ...Centrifuge CM 50 User s manual ISO 9001 2008...

Page 2: ...and usage 3 Preparation to work 4 Control description 4 Operation 5 Comparison table of RPM to RCF and vice versa 6 Troubleshooting 7 Emergency opening of centrifuge lid 7 Safety precautions 8 Disinfe...

Page 3: ...r RCF units blocking of lid during work sound alarm and signals sleep mode and automatic lid unblocking at the end of work Built in temperature and imbalance sensors protect the unit from overheating...

Page 4: ...nto the grooves of the rotor then tighten the rotor nut with special key Usage always load the rotor symmetrically By minimizing the difference of weight between the filled test tubes you reduce the w...

Page 5: ...e of faults or damage the centrifuge can be considered ready to work ATTENTION In the presence of faults and damage do not use the centrifuge without consulting a specialist ATTENTION Network socket m...

Page 6: ...trifuge stops alarms sounds and lid unblocks Centrifuge 50 supports a special PUSH mode to make your work more effective and a sleep mode to minimize power consumption PUSH mode is in the speed range...

Page 7: ...6 Comparison table of rounds per minute RPM to relative centrifugal force RCF and vice versa Operation...

Page 8: ...is shown on the figure Er_td Er_bd Er_db Er_OP No display Error Overheating Rotor is blocked Engine failure Total imbalance of test tubes is more than 1 gram Lid not closed No main power connection P...

Page 9: ...d transportation runs accordingly to regulations applied to this way of transport Equipment should be stored in original manufacturers packaging in dry room with humidity not more then 80 and temperat...

Page 10: ...ture_________________________________ Place for stamp Certificate of sale Organization _______________________________________________ Address __________________________________________________ Phone_...

Page 11: ......

Reviews: