Trouble Shooting
© Elma Schmidbauer GmbH
BA/US/MF/1202
19
11
Trouble Shooting
In case of damage please contact the manufacturer.
Repair and maintenance works on the connected and
opened unit must be carried out by
specialised and
ELMA-
authorised personnel
only.
Separate the unit from the mains supply before opening it.
Faulty parts must be replaced by
original spare parts
only
to guarantee the valid safety norms.
The
knob shells
of the operating elements can be replaced
from outside without opening the unit.
Operating elements
(timer; control; switch): to replace, remove
the four screws from the front panel and carefully take off the
front panel.
Possible causes:
no voltage supply to the unit
mains cable damaged (interrupt)
frequency selector switch in 0 position
timer in off position
HF cable not connected
HF cable / plug damaged
ultrasonic power module(s) faulty
fault in the transducer system
In case of more than one module and transducer system
connections they can be interchanged to check their functioning
and to find the fault
faulty remote control
faulty SPC
In order to eliminate any fault on these components, remove
them for the functional check-up of the generator.
Check-up of the transducer system by means of an
ohmmeter.
Damage on the
operating elements
No ultrasound