Operating instructions 

 

 

                       

 

 

 

 

 

 

Elmasonic MF  

Ultrasonic-generators 

 

 english 

 

 

 

MF 200 

– MF 1000 

MF 2000 

– MF 4000  

MF 2000 

– MF 4000 

Summary of Contents for Elmasonic MF 1000

Page 1: ...7 D 78224 Singen Phone 49 0 7731 882 0 Fax 49 0 7731 882 266 info elma ultrasonic com www elma ultrasonic com document ref 106 4799 Operating instructions Elmasonic MF Ultrasonic generators english MF...

Page 2: ...etting of frequency 9 6 Remote Control Optional 10 7 Cleaning Process and Cleaning Chemicals 12 7 1 Ultrasonic cleaning effect 12 7 2 Cleaning chemicals 12 7 3 Degassing of the liquid 12 7 4 Cleaning...

Page 3: ...dry Keep working area and operating elements clean and dry to prevent any electricity related accidents The transducer elements and the heating if available must be completely covered in liquid Dry ru...

Page 4: ...rasonic energy into the cleaning bath at a high efficiency rate This modern ELMA ultrasound equipment consists of a high performance generator secured against open circuit and short circuit and the co...

Page 5: ...back of the unit MF 200 MF 1000 3 2 Available connections on the back of the unit MF 2000 MF 4000 Mains Input Malfuncion Indicator for SPC Remote Control SPC Ventilator HF connection Ventilator 2x HF...

Page 6: ...the operation e g by means of a level control device The level control prevents operation of the unit with an insufficient quantity of liquid in the tank Experience has proved the release of the gener...

Page 7: ...a unit with heating produces high temperatures in the tank Caution Risk of burning Do not reach inside the tank during ultrasonic operation Caution Cellular membranes will be damaged by prolonged exp...

Page 8: ...to 15 minutes the unit can be operated by means of the automatic timer The required cleaning period is selected by turning the knob When the selected period is terminated the unit switches off automa...

Page 9: ...asound transformer This optimises the performance in the distributed maxima The set power is interrupted for a short period so that the bubbles are not retained by the ultrasonic forces The degassing...

Page 10: ...control please see also chapter Operation Connection of remote control to generator please see chapter Connections Order Numbers Remote Control Frequency MF 600 1000 MF 2000 4000 25 45 KHz 131000268 1...

Page 11: ...of remote control Operating options Timer power control selection of ultrasound mode sweep normal degas selection of frequency For detailed information of the functions please see chapter Operation Fo...

Page 12: ...huric acid or nitric acid ELMA has a wide range of cleaning chemicals on offer developed and made in our own laboratory The application of these chemicals help to prolong the service life of the ultra...

Page 13: ...items The optimum temperature for your cleaning task the recommended concentration of the cleaning chemical and the required cleaning period are indicated on the label of the used elma clean chemical...

Page 14: ...ings in accordance with the EC regulations E or R 1 R 2 or R 3 for combustible materials or F F or R 10 R 11 or R 12 for inflammable materials may not be placed in the stainless steel tank of the ultr...

Page 15: ...actured by ELMA Page 9 Cleaning Agents Areas of Use Examples list not complete treatment with hydrochloric acid or hydrofluoric acid or with salts of acid solutions removal of fluxing agents containin...

Page 16: ...s transducer plates During operation the transducers are exposed to extreme loads and must be checked for any damages at regular intervals monthly Possible marks and damages due to wear and tear are t...

Page 17: ...anti liming media on citric acid or phosphorous acid basis 10 Technical Data Approval and Norms allowed ambient temperature 5 C to 40 C allowed relative humidity of air up to 30 C 80 allowed relative...

Page 18: ...Technical Data Approval and Norms 18 BA US MF 1202 Elma Schmidbauer GmbH Technical Data...

Page 19: ...ol switch to replace remove the four screws from the front panel and carefully take off the front panel Possible causes no voltage supply to the unit mains cable damaged interrupt frequency selector s...

Page 20: ...r is indicated by 2 3 or 4 green lamps in the display small LED G1 G4 depending on the type of generator MF 2000 3000 4000 If one module of the generator is faulty the GREEN lamp turns RED The generat...

Page 21: ...e generator is switched off timer at 0 position If one module breaks down during operation the fault is indicated as described above For switching off one of the modules through the status indication...

Page 22: ...kpl 25 45KHz 100V Generator cpl 25 45 KHz 100V G n rateur cpl 25 45KHz 100V 1301000289 Generator kpl 25 45KHz 200V 208V Generator cpl 25 45 KHz 200V 208V G n rateur cpl 25 45KHz 200V 208V 1301000319...

Page 23: ...000415 2000000415 Isolierschlauch MOS FET Isolating tube MOS FET Tube isolant MOS FET 8000003564 8000003564 8000003564 2000003564 Entst rfilter 230V RFI Filter 230V Filtre antiparasite 230V 2000001098...

Page 24: ...00005114 2000005114 2000005114 Drehknopf f r Zeitschaltuhr Knob for timer Bouton pour minuterie 2000000947 2000000947 2000000947 2000000947 Potentiometer Power Potentiometer Power Potentiom tre Power...

Page 25: ...t can also be returned to the manufacturer 15 Manufacturer Contact Elma Schmidbauer GmbH Kolpingstr 1 7 D 78224 Singen Phone 49 0 7731 882 0 Fax 49 0 7731 882 266 Homepage www elma ultrasonic com Tech...

Page 26: ...Notes 26 BA US MF 1202 Elma Schmidbauer GmbH Notes...

Page 27: ...Notes Elma Schmidbauer GmbH BA US MF 1202 27 Notes...

Page 28: ...Notes 28 BA US MF 1202 Elma Schmidbauer GmbH Notes...