Technical specifications

Elmadry TD

TD 30

TD 120

TD 300

Mechanical data

Tank internal
dimensions W/D/H
(mm)

240/137/100

300/240/200

505/300/200

Unit external
dimensions W/D/H
(mm)

300/186/219

362/285/322

568/347/322

Weight (kg)

3,6

6,1

10,2

Maximum load (kg)

1

7

12

Mesh width of wire
mesh (mm)

2

2

2

Wire diameter of
wire mesh (mm)

0,63

0,63

0,63

Material (tank/
enclosure)

Chrome steel

Chrome steel

Chrome steel

Electrical data

Mains voltage
(VAC) (50/60 Hz)

100-120 / 220-240

100-120 / 220-240

100-120 / 220-240

Heating power at
110 and 230 V (W)

480

550

1100

Power at 110 and
230 V (W)

500

570

1140

Max. temperature
(°C)

70

70

70

Heating time (min.)

~15

~15

~15

Air output max. (m

3

/

h)

160

160

320

Accessories

Suitable basket
sizes

1x 30

2x 30/40

1x 100/120

3x 60

1x 60 + 1x 100/120

2x 100/120

1x 150/300

Summary of Contents for Elmadry TD

Page 1: ...Instruction Elmadry TD Hot air dryer english Version 0 Elma Schmidbauer GmbH Gottlieb Daimler Str 17 D 78224 Singen Tel 49 7731 882 0 Fax 49 7731 882 266 info elma ultrasonic com www elma ultrasonic...

Page 2: ...phics plans etc may be reproduced or distributed without our written permission nor may it be used in an unauthorised manner for the purpose of competition nor passed to third parties or made accessib...

Page 3: ...symbols used 9 5 3 Safety information on the unit 9 5 4 Safety information on using the unit 10 6 Product contents 11 6 1 Optional accessories 11 7 Unit description 12 8 Unpacking and mounting the uni...

Page 4: ...d This symbol denotes lists This symbol denotes requirements Numbers with a dot denote actions This symbol denotes individual actions This symbol denotes interim results This symbol denotes the result...

Page 5: ...aning It is suitable for drying clock movements metal strips clock parts large silverware and housings Do NOT dry objects with solvents on them The unit is only intended for use in commercial applicat...

Page 6: ...ate using the rocker switch with three positions OFF HOT hot air COLD cold air Dust filter replaceable prevents dust in the ambient air from re contaminating the item Cooling unit plug in connection o...

Page 7: ...g 1 7 12 Mesh width of wire mesh mm 2 2 2 Wire diameter of wire mesh mm 0 63 0 63 0 63 Material tank enclosure Chrome steel Chrome steel Chrome steel Electrical data Mains voltage VAC 50 60 Hz 100 120...

Page 8: ...disregarded Measures to prevent the danger Signal word Meaning Consequences if disregarded DANGER Immediate danger Death or serious injury WARNING Possible dangerous situation Death or serious injury...

Page 9: ...Hot surfaces Symbols in the manual Risk of electric shock Risk of explosion due to oxidising substances Note about damage to unit or tip Disposal information 5 3 Safety information on the unit CAUTION...

Page 10: ...ot place the unit near flammable substances Do not dry objects with solvent adhered to them alcohol benzine etc DANGER Liquid ingress Electric shock Do not immerse unit in water or rinse it under runn...

Page 11: ...lustration 1 Elmadry TD product contents 1 Lid 2 Network cable 3 Elmadry TD 4 Manual not pictured 6 1 Optional accessories Silicon mat to place on the bottom of the tank prevents scratching of parts S...

Page 12: ...D 7 Unit description 1 2 3 4 5 Illustration 2 Front of Elmadry TD 300 1 Lid to cover the unit or as a place to temporarily stand a stainless steel basket 2 Handle for transporting 3 Foot 4 Rocker swit...

Page 13: ...24 1 2 3 Illustration 3 Rear of Elmadry TD 300 1 Mains connection for connecting with the network cable 2 Name plate shows the technical specifications for the unit 3 Dust filter replaceable prevents...

Page 14: ...n 4 Dryer tank of Elmadry TD 300 1 Tubular heater for heating the air 2 Wire mesh for protection of the blower system 3 Blower generates a stream of air to dry the parts 4 Over temperature fuse switch...

Page 15: ...packaging can be sent back to the manufacturer NOTICE Permissible ambient conditions Incorrect ambient conditions lead to damage to the unit Permissible ambient temperature 5 C to 40 C 41 104 F Permis...

Page 16: ...socket In emergencies pull the plug out of the socket or the unit Keep the supporting surface and the enclosure dry The information on the name plate must match the electrical operating conditions par...

Page 17: ...re than 70 NOTICE Too much water inside the unit leads to damage Allow excess water to drip off the basket and the objects before drying NOTICE To prevent stain formation rinse the objects thoroughly...

Page 18: ...ot air set the rocker switch to HOT 5 To dry with cold air or cool the objects set the rocker switch to COLD 6 To switch off the unit set the rocker switch to OFF 7 After drying allow the objects to c...

Page 19: ...The unit has cooled down 1 Clean any dust residue on the wire mesh using suction 2 Remove other residue on the wire mesh with a damp cloth 3 Remove any residue on the tank with a damp but not wet clot...

Page 20: ...bauer GmbH 20 24 Elmadry TD 3 Remove the dust filter 4 Insert a new dust filter or clean and replace the old dust filter 5 Replace the cover 6 Turn the unit back to the upright position The dust filte...

Page 21: ...faulty or blocked with dirt Contact the manufacturer The objects are dirty after the drying process Soiled dust filter Change or clean the dust filter Heating and or blowing do not work Unit faulty Co...