Elma 718
side 21
TX
RX
GND
Low Battery Condition
When the battery voltage is under proper operation requirement, the symbol will show
on the LCD and the battery need to be replaced with new one.
Calibration Point:
input Adjust
VR
tolerance
0 °C
VR1
± 0.1 °C
190 °C
VR2
± 0.1 °C
1000 °C
VR3
± 1 °C
1900 °F
VR4
± 1 °F
Digital Output:
The Digital Output is a 9600bps N 81 serial interface.
The RX is a 5V normal high input port.
The TX is a 5V normal high output port.
V. Setup TestLink SE-309
—
RS232 interface software:
•
The TestLink package contains:
1. 80mm CD.
2. Custom designed RS232 cable for TestLink.
•
System Required:
Windows 95 or Windows 98 or Windows NT 4.0 above.
•
Minimum Hardware Required:
PC or NoteBook with Pentium 90MHz or higher, 32 MB
RAM ; At least 5 MB byte hard disk space available to install TestLink.
Recommended resolution 800X600.
P.S
Normally, performing offset Calibration with thermal
stabled ice water through VR1 will give a very good
calibration result.