Elma VT-8204
side 36
How to send the data out from the meter
1. Before sending data out from the meter, exit the "Hold function " and the " Record " function.
2. Press the " Send Button " ( 3-5, Fig. 1 ) at least 2 seconds until display show " r-232 ", then
release the
button.
3. Push the " Send Button " ( 3-5, Fig.1) once, display will count down, at the same the storage
data logger data will send out the meter from the " RS-232 Output Terminal " ( 3-19, Fig. 1 ).
4. If intend load the data to the computer, it should connect the RS232 cable ( optional, model :
UPCB-02) or USB cable ( optional, model : USB-01 ) and apply the Data Logger software (
optional, Model : SW-DL2005 ).
Auto power off
Auto power off Is default. If the user intend to disable the " Auto Power off "function.
Note :
During execute the record function, the auto power function will disable too.
Advanced setting procedure
1. Power off the meter, first press the the " Hold button " ( 3-3, Fig. 1 ) continuously, then press
the " Power button " ( 3-2, Fig. 1 ) once. Release from " Hold button ".
2. One by one to press the " Hold button " ( 3-3, Fig. 1 ) once a while to select the five four
function and display will show flashing text with as :
OFF.......
Auto power On/Off management
SEC......
Change the data logger sampling time
Cnt........
To show the balance data numbers in the memory
CLr........
Clear the existing saving data from the memory
ESC.......
Escape the advanced setting function
3. Auto power On/Off
a. Use the " Hold button " to select the main function to " OFF ".
b. Press the " Function button " (3-5, Fig.1) or " Unit button " (3-6, Fig.1) to select "1" or "0".
* If the display value show " 0 ", it will disable the Auto Power Off function.
* If the display value show " 1 ", it will execute the Auto PowerOff function.
c. After select the desiring value ( 1 or 0 ), press the "
Enter button " ( 3-4, Fig. 1 ) to save the data with default.
4. Change the data logger sampling time
a. Use the " Hold Button " to select the main function to " SEC ".
b. Press the " Function button " ( 3-5, Fig. 1) or " Unit button " ( 3-6, Fig. 1) to select the data
logger sampling time to 0, 1, 2, 10, 30, 60, 600, 1800, 3600 seconds.
c. After the sampling time value is determined, press the " Enter button " ( 3-4, Fig. 1 ) to
save the sampling time with default.
5. To show the balance data numbers in the memory. Use the " Hold button " to select the
main function to " Cnt ". In the same time display will show the balance data point that exist
into the memory ( allow memorize data no. ).
6. Clear the existing saving data from the memory
a. Use the " Hold button " to select the main function to " CLr ".
b. Press the " Function button " ( 3-5, Fig. 1) or " Unit button " ( 3-6, Fig. 1) to select " 1 " or "
0 ".
* If the display value show " 0 " , it will be not to clear the memory.
* If the display value show " 1 " will execute the memory clear function..
c. After select the desiring value ( 1 or 0 ), press the "Enter button " ( 3-4, Fig. 1 ) to save
the data with default.
7. ESC
a. Use the " Hold button " to select the main function to " ESC ".
b. Press the " ESC button " ( 3-3, Fig. 1 ) will escape the above advanced setting function.