Manual or Auto-ranging 

Press the RANGE button momentarily to select manual-ranging, and the meter will remain in the 
range it was in, the LCD 

 turns off.  Press the button momentarily again to step through the 

ranges.  Press and hold the button for 1 second or more to resume auto-ranging. 

Note:

 Manual ranging feature is not available in Hz, 

, RPM, ms, DWELL & Duty functions. 

Set Beeper Off 

Press the RANGE button while turning the meter on to temporarily disable the Beeper feature. 
Turn the rotary switch OFF and then back on to resume. 

Beep-

Jack™ Input Warning 

The meter beeps as well as dis

plays “InEr” to warn the user against possible damage to the meter 

due to improper connections to the 

A, mA, or A input jacks when other function (like voltage 

function) is selected. 

Auto-Power-Off (APO) 

The Auto-Power-Off (APO) mode turns the meter off automatically to extend battery life after 
approximately 34 minutes of no rotary switch or push button operations. To wake up the meter 
from APO, press the SELECT button momentarily or turn the rotary switch OFF and then back on. 
Always turn the rotary switch to the OFF position when the meter is not in use 

Disabling Auto-Power-Off 

Press the SELECT button while turning the meter on to temporarily disable the Auto-Power-Off 
(APO) feature. Turn the rotary switch OFF and then back on to resume. 

Maintenance 

WARNING 

To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the 
input jacks and turn OFF the meter before opening the case. Do not operate with open case. 
Install only the same type of fuse or equivalent 

Calibration 

Periodic calibration at intervals of one year is recommended to maintain meter accuracy. Accuracy 
is specified for a period of one year after calibration. If self-diagnostic message 

“C_Er” is being 

displayed while powering on, some meter ranges might be largely out of specifications. To avoid 
mis-leading measurements, stop using the meter and send it for re-calibration. Refer to the 
LIMITED WARRANTY section for obtaining warranty or repairing service. 

 

Cleaning and Storage 

Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. 
If the meter is not to be used for periods of longer than 60 days, remove the battery and store it 
separately  

Trouble Shooting 

If the instrument fails to operate, check battery, fuses, leads, etc., and replace as necessary. 
Double check operating procedure as described in this user’s manual 
 
If the instrument voltage-resistance input terminal has subjected to high voltage transient (caused 
by lightning or switching surge to the system) by accident or abnormal conditions of operation, the 
series fusible resistors will be blown off (become high impedance) like fuses to protect the user 
and the instrument. Most measuring functions through this terminal will then be open circuit. The 
series fusible resistors and the spark gaps should then be replaced by qualified technician. Refer 
to the LIMITED WARRANTY section for obtaining warranty or repairing service. 

Summary of Contents for Elma 319s

Page 1: ...Elma 319s Dansk norsk manual Side 5 12 Svenska manual Sida 14 22 English user manual Page 23 32 EAN 5706445410385...

Page 2: ...er 8 A mA og A funktioner 9 IG RPM eller M funktion 9 DWELL Duty funktioner 10 ms FUEL INJECTION DETECTOR funktion 10 IP RPM eller M funktion 11 Baggrundsbelyst LCD display 11 Hold funktion 11 Manuel...

Page 3: ...Kapacitans Diodtest 17 DCmV ACmV Temperatur oC oF 17 A mA och A Str m 18 IG RPM DWELL DUTY ms 18 DWELL Duty 19 ms FUEL INJECTION DETECTOR 19 IP RPM eller M 20 Bakgrundsbelyst LCD display 20 Hold 20 Ma...

Page 4: ...ns 26 A mA and A Current functions 27 IG RPM DWELL DUTY ms functions 27 IG RPM or M function 27 DWELL Duty functions 28 ms FUEL INJECTION DETECTOR function 28 IP RPM or M function 29 Backlighted LCD d...

Page 5: ...Kat IV 300V AC DC M lekategori II defineres som m ling i stikkontakter og udtag mere end 10m fra kategori III 20m fra kategori IV og m ling i brugsgenstande WARNING For at undg risiko for elektrisk st...

Page 6: ...l CENELEC lavsp ndingsdirektivet 2006 95 IEC og elektromagnetisk kompatibilitetsdirektivet 2004 108 EC Produktbeskrivelse 1 3 5 6 digits 6000 counts display 2 Trykknapper for specielle funktioner 3 Dr...

Page 7: ...AUTION B de f r og efter en sp ndingsm ling b r man kontrollere multimetret p en kendt sp nding for kontrol af om multimetret m ler korrekt DCV ACV Frekvensfunktioner Tryk p SELECT knappen op til fler...

Page 8: ...r og funktion af switche Et kontinuerligt bip indikerer at ledningen er intakt Normal sp ndingsfald forward biased for en god silikonediode l gger imellem 0 400V til 0 900V En afl sning h jere end det...

Page 9: ...rne hukommelse IG RPM eller M funktion Tryk p knappen flere gange for at v lge RPM for 4 cylinder RPM for 2 cylinder og DIS og RPM M for speciel 2 cylinder 2 takts motor Fra fabrikkens side er antalle...

Page 10: ...trigger niveau Hvis omdrejnings RPM afl sningerne bliver ustabile skal der v lges en lavere f lsomhed for at undg elektrisk st j Hvis afl sningen viser 0 v lg da en h jere f lsomhed Antal af mark rpu...

Page 11: ...nktion Hold funktionen fryser afl sningen i displayet Tryk p HOLD knappen for at aktivere hold funktionen Manuel eller auto omr de Tryk p RANGE knappen flere gange for at v lge manuel omr de og instru...

Page 12: ...1 r efter kalibreringsdato Hvis Selvdiagnose beskeden C_Er vises n r der t ndes for Elma 319s kan det ske at instrumentet er ude af specifikationer For at undg fejlm linger b r instrumentet sendes ti...

Page 13: ...V m er ikke specificeret Overbelastnings Beskyttelse A mA 0 4A 1000V DC AC rms IR 30kA Type F A 11A 1000V DC AC rms IR 20kA Type F V 1100V DC AC rms mV Ohm others 1000V DC AC rms Lavt batteri Under ca...

Page 14: ...ori II CAT II r f r m tningar utf rda p kretsar anslutna direct till en l gvoltsinstallation Exempel r m tningar p hush llsapparater portabla verktyg och liknande VARNING F r att reducera risken f r e...

Page 15: ...ering S kring AC V xelstr m DC Likstr m CENELEC DIREKTIV Instrumentet verensst mmer med CENELEC L gvoltsdirektiv 2006 95 EC och Elektromagnetiska kompabilitetsdirektivet 2004 108 EC PRODUKTBESKRIVNING...

Page 16: ...lande till RMS v rdena ANV NDNING VARSAMHET F re och efter m tningar med farliga sp nningar testa sp nningsfunktionen p en k nd sp nning f r att konstatera korrekt function DCV ACV Frekvensfunktionern...

Page 17: ...ledare anslutningar samt function hos brytare En kontinuerlig summersignal indikerar kontakt Normal framsp nning f r en ok silikondiod r mellan 0 400V och 0 900V En svl sning h gra n det indikerar en...

Page 18: ...gning IG RPM eller M Tryck p knappen kort f r att v lja mellan RPM f r 4 takt RPM f r 2 takt och DIS samt RPM M f r special 2 takts waste ignition engine Antalet cylindrar r f rinst llt p 4 cylindrar...

Page 19: ...trycka p TRIGGER knappen I mer n en sekund i Duty funktionen f r avancerade applikationer Trigger niv n r f rinst lld p 3 Tryck p Level RANGE knappen kort f r att v lja en annan trigger niv Om DWELL e...

Page 20: ...ningen visar noll v lj en h gre k nslighet Numret p Bargrafens pekare anv nds f r att indikera vald k nslighet Bakgrundsbelyst LCD display Tryck p SELECT knappen I mer n en sekund f r att aktivera bak...

Page 21: ...lla instrumentets noggrannhet Noggrannheten r specificerad f r en period om ett r efter kalibrering Om meddelandet C_Er visas n r instrumentet sl s p kan det betyda att visa m tomr den r utanf r spec...

Page 22: ...S kring F2 for A mA 0 4A 1000V AC DC IR 30kA Type F 6 x 32 mm S kring F1 for A 11A 1000V AC DC IR 20kA Type F 10 x 38 mm Batteri oh s kringsbyte Lossa skruven p baksidan Lyft av luckan Byt batterier...

Page 23: ...surements performed on circuits directly connected to the low voltage installation Examples are measurements on household appliances portable tools and similar equipment WARNING To reduce the risk of...

Page 24: ...Fuse AC Alternating Current DC Direct Current Cenelec directives The instruments conform to CENELEC Low voltage directive 2006 95 EC and Electromagnetic compatibility directive 2004 108 EC Product des...

Page 25: ...s Operation CAUTION Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to determine proper meter functioning DCV ACV Line Frequency functi...

Page 26: ...one indicates a complete wire Normal forward voltage drop forward biased for a good silicon diode is between 0 400V to 0 900V A reading higher than that indicates a leaky diode defective A zero readin...

Page 27: ...ons in sequence Last selection will be saved as power up default for repeat measurement convenience IG RPM or M function Press button momentarily to select through RPM for 4 stroke RPM for 2 stroke an...

Page 28: ...angle has been adjusted Press SELECT button momentarily to display DWELL reading in terms of percentage if required Positive negative trigger slopes are selectable through pressing TRIGGER button for...

Page 29: ...hrottle Body Injectors TBI which operate with twin pulses IP RPM or M function Press button momentarily to select through RPM for 4 stroke RPM for 2 stroke and DIS and RPM M for special 2 stroke waste...

Page 30: ...circuit remove the test leads from the input jacks and turn OFF the meter before opening the case Do not operate with open case Install only the same type of fuse or equivalent Calibration Periodic c...

Page 31: ...C or 28oC 40oC or otherwise specified Sensing Average sensing Pollution Degree 2 Safety Double insulation per IEC UL EN61010 1 Ed 3 0 IEC EN61010 2 030 Ed 1 0 IEC EN61010 2 033 Ed 1 0 IEC UL EN61010 0...

Page 32: ...BKP60 banana plug type K thermocouple BP300 Inductive pickup clip Optional purchase accessories Magnetic hanger BMH 01 BKB32 banana plug to type K socket plug adaptor Electrical Specification Accurac...

Page 33: ...5d 60 00mA 1 2 2 5d 2 5 mV mA 600 0mA 1 2 0 5d 6 000A 2 2 5d 0 03V A 10 00A 2 1 2 5d 1 Ranges for Model 319s only 2 6A continuous 6A to 10A for 30 sec max with 5 minutes cool down interval Temperature...

Page 34: ...s cyl Nine selectable Cylinders 1 2 3 4 5 6 8 10 12 Four selectable trigger levels Sensitivity Level 1 0 8V typically Level 2 1 85V typically Level 3 3 75V typically Level 4 6V typically Fuel injectio...

Page 35: ...v 181101...

Reviews: