Elma 117 - 118
Page 48
example. If the reading shows zero due to insufficient signal levels, select Line-Level Hz
function BEFORE making measurements (at AC 6V range) will set the highest sensitivity.
AutoCheck
TM
mode
Set the slide-switch function selector to the
position. This innovative AutoCheck
TM
feature
automatically selects measurement function of DCV, ACV or Resistance (
Ω
) based on the input via
the test leads.
•
With no input, the meter displays “Auto” when it is ready.
•
With no voltage signal but a resistance below 6M
Ω
is present, the meter displays the
resistance value. When below 25
Ω
(0.025k
Ω
) is present, the meter further gives a
continuity beep tone.
•
When a signal above the threshold of +1.5 VDC, -1 VDC or 2 VAC up to the rated 600V is
present, the meter displays the voltage value in appropriate DC or AC, whichever larger in
peak magnitude.
Overload-Alert Feature:
When above rated 600V is present, the meter displays “OL” with a
warning beep tone for over-range indication. Disconnect the test leads from the signal
immediately to avoid hazards.